1.
目的 对科华ELISA和雅培i2000sr检测系统4种感染性疾病8项血清学标志物的性能进行评价和比较。方法 采用美国临床和实验室标准协会(clinical and laboratory standards institute,CLSI)评价方案(evaluation protocols,EP)评价两种检测系统性能。评价内容包括:准确度、检出限、一致性、精密度、线性范围。结果 两种检测系统的准确度均为100%。HBsAg,HBsAb,HBeAg,HBeAb,HBcAb,Anti-HCV,梅毒特异性抗体及(Anti-TP)Anti-HIV的检出限科华分别为0.08IU/ml,10mIU/ml,0.8NCU/ml,4.4NCU/ml,3.2NCU/ml,0.8NCU/ml,20 mIU/ml和0.5 NCU/ml; 雅培分别为0.05 IU/ml,10 mIU/ml,0.23 NCU/ml,1 NCU/ml,1.8 NCU/ml,0.45 NCU/ml,20 mIU/ml和2.88 NCU/ml。一致性验证:两种系统方法学比对Kappa值分布于0.8~0.9,定性项目仪器间比对Kappa值均为1, 定量项目(HBsAg,HBsAb)仪器间比对观察误差<1/2TEa,浓度误差指数绝对值<1。精密度评价:批内变异系数(CV)范围为1.10%~7.01%,室内CV范围为1.90%~10.03%。雅培i2000sr HBsAg,HBsAb线性范围分别为0.73~245IU/ml和4.21~960.89mIU/ml。结论 科华ELISA和雅培i2000sr准确度、检出限、一致性、精密度和线性范围与厂家声明一致并能满足实验室要求。两种系统HBsAg,HBsAb,Anti-TP检出限接近,但HBeAg,HBeAb,HBcAb,Anti-HCV和Anti-HIV检出限二者存在较大差异。  相似文献   

2.
目的对珠海丽珠公司生产的乙型病毒性肝炎(简称乙肝)酶联免疫吸附试验(ELISA)试剂盒(以下简称"珠海丽珠试剂盒")进行性能验证,为实验室选用合适的方法和试剂提供依据。方法参考美国临床实验室标准化研究所EP15-A2标准,采用重复性代替精密度,以乙肝标志物的阳性、阴性符合率和一致率分别作为批内重复性精密度和批间重现性精密度评价指标;选取2015-2016年卫生部临检中心感染性血清学标志物A室间质评物进行准确性验证;将已知浓度的高值质控品用阴性患者新鲜血清作梯度稀释后进行检出限验证;以罗氏电化学发光免疫分析法(ECLIA)检测结果为标准,对灵敏度、特异度及总符合率进行验证,同时,对ELISA、胶体金免疫层析法(GICT)和ECLIA检测乙型肝炎表面抗原(HBsAg)的灵敏度、特异度等进行比较。结果珠海丽珠试剂盒的批内重现性精密度符合率、批间重现性精密度一致率、准确度符合率均为100%。HBsAg、乙型肝炎表面抗体(HBsAb)、乙型肝炎e抗原(HBeAg)、乙型肝炎e抗体(HBeAb)、乙型肝炎核心抗体(HBcAb)的检出限分别为1.33IU/mL、30.00mIU/mL、2.00NCU/mL、4.00NCU/mL、0.44NCU/mL。珠海丽珠试剂盒检测HBsAg、HBsAb、HBeAg、HBeAb、HBcAb的灵敏度、特异度、总符合率及约登指数均达到较高的水平。ELISA检测HBsAg的灵敏度优于GICT。结论珠海丽珠试剂盒具有较高的精密度、灵敏度及特异度,能够满足临床检测的质量要求。  相似文献   

3.
目的验证乙型肝炎(简称乙肝)5项酶联免疫吸附试验(ELISA)定性检测的分析性能,改进和优化试验方案,为实验室选用合适的方法和试剂提供依据。方法通过梯度稀释定值样品测定检出限;依据美国临床实验室标准化协会(CLSI)EP12-A2文件进行临界值验证;选取临界值样品评价分析精密度(包括重复性和中间精密度);通过与室间质评检测结果及雅培i4000SR化学发光仪比对评估符合率。结果乙肝表面抗原(HBsAg)、乙肝表面抗体(HBsAb)、乙肝e抗原(HBeAg)、乙肝e抗体(HBeAb)及乙肝核心抗原(HBcAb)的最低检出限分别为:0.2IU/mL、20mIU/mL、1NCU/mL、0.75NCU/mL、0.05NCU/mL。临界值(C50)±20%的浓度范围包含了多次重复试验阳性率5%~95%的浓度点区间。批内变异系数(CV)≤15%,夹心法批间CV≤25%,竞争法批间CV≤35%。准确度及与雅培i4000SR比对的阴、阳性符合率均大于或等于95%,其一致性检验κ≥0.75。结论本实验室乙肝5项ELISA检测方法性能符合检测分析要求,检测质量可满足临床需求。  相似文献   

4.
目的对磁微粒化学发光法检测乙型肝炎6项血清标志物[乙型肝炎表面抗原(HBs Ag)、乙型肝炎表面抗体(HBs Ab)、乙型肝炎e抗原(HBe Ag)、乙型肝炎e抗体(HBe Ab)、乙型肝炎核心抗体(HBc Ab)、乙型肝炎前S1抗原(PreS1Ag)]的关键性能指标进行验证。方法参照相关标准,对磁微粒化学发光法检测乙型肝炎6项血清标志物的精密度、线性范围、临床可报告范围、参考区间和检出限进行验证,并比较检测系统使用前、后的样本周转时间(TAT)。结果磁微粒化学发光法检测乙型肝炎6项血清标志物的实验室内变异系数(CV)、批内s(重复性不精密度)和实验室内不精密度均低于标准规定范围。HBsAg、HBsAb、HBeAg、HBeAb、HBcAb、PreS1Ag的临床可报告范围分别为0.01~400.00 IU/mL、0.57~1 948.90 mIU/mL、0.01~203.36 PEIU/mL、0.02~17.24 PEIU/mL、0.20~79.23 PEIU/mL和0.58~246.13 AU/mL,与厂商提供的临床可报告范围接近。参考区间验证通过,可使用厂商提供的参考区间。检测5份与厂家声明的检出限(LOD)水平相当的样本,所有检测结果均大于厂家声明的空白限(LOB),LOD验证通过。与使用免疫流水线前的TAT比较,使用免疫流水线后的TAT明显缩短(P0.000 1)。结论磁微粒化学发光法检测乙型肝炎6项血清标志物的关键性能指标均符合临床要求。使用免疫流水线后TAT明显缩短。  相似文献   

5.
乡镇医院乙型肝炎病毒血清标志物室间质评分析   总被引:1,自引:0,他引:1  
目的了解乡镇医院乙型肝炎病毒血清标志物检测结果准确度。方法制备乙型肝炎病毒标志物血清质控品,每批均含低浓度样本,乙型肝炎病毒表面抗原(HBsAg)约0.5~1ng/mL,乙型肝炎病毒表面抗体(抗-HBs)约20-30mIU/mL,乙型肝炎病毒e抗原(HBeAg)约2NCU/mL,乙型肝炎病毒e抗体(抗-HBe)约4NCU/mL,乙型肝炎病毒核心抗体(抗-HBc)约2NCU/mL。每批5个样本,共分3批,发放各实验室检测。结果与预期结果比较,参加实验室第1批次正确率为78%,第2批次和第3批次正确率皆为84%。结论乡镇医院乙型肝炎病毒血清标志物检测结果准确度偏低,急需改进和提高。  相似文献   

6.
化学发光法定量检测乙型肝炎病毒标志物临床应用评价   总被引:2,自引:1,他引:1  
目的对化学发光法定量检测乙型肝炎病毒标志物的临床应用进行评价。方法用标准参考血清和定值质控血清,对化学发光法定量检测乙型肝炎病毒标志物进行评价,评价内容为符合率、分析灵敏度、分析特异性、批内和批间精密度、最低检出限等指标。结果HBsAg符合率为97.5%,其他四项符合率大于92%;HBsAg分析灵敏度100%、分析特异性96.15%;五项的批内和批间精密度均小于15%;最低检出限分别为HBsAg0.06IU/mL、HBsAb0.5mIU/mL、HBeAg0.05NCU/mL、HBeAb0.5NCU/mL、HBcAb0.5NCU/mL。结论化学发光法定量检测乙型肝炎病毒标志物各项性能指标能满足临床要求,但操作需注意标本质量。  相似文献   

7.
目的 建立并评价化学发光免疫法检测甲胎蛋白(AFP)的空白限(LoB)、检出限(LoD)和定量检测限(LoQ)。 方法 参考CLSI EP17-A文件,将AFP空白样品和系列低浓度样品在Bayer Centaur 240化学发光免疫检测系统上进行检测,依据数据的分布规律,采用相应的统计学方法,确定血清AFP的LoB、LoD和LoQ,同时用传统方法建立AFP的检测低限(LLD)、生物检测限(BLD)和功能灵敏度(FS)。 结果 用EP17-A方案建立的LoB为0.80 ng/ml,LoD为2.60 ng/ml,LoQ为3.30 ng/ml;应用常规方法建立的LLD为1.00 ng/ml,BLD为2.20 ng/ml,FS为2.20 ng/ml。 结论 应用两种方法建立的LoB和LLD均低于仪器说明书声明的灵敏度性能,Bayer Centaur 240化学发光免疫系统检测AFP的分析灵敏度得到验证确认;实验室在验证仪器声明的LoB时,应同时建立检测系统的LoD和LoQ,以便为临床提供更为可靠的检验结果。两种评价方案均具有实用性,国内实验室可视具体情况选择不同的评价方案。  相似文献   

8.
当体外诊断试剂厂商仅按医药行业标准对某个定性检测项目声明了空白限(LoB)时,临床实验室应完成该项目检出限(LoD)的性能验证。文章对多个医学相关个行业标准中LoD、LoB的定义进行分析,并指出其在各文件中的异同。同时将定性项目中C50浓度、C95浓度与定量项目中LoD、LoB的相互联系进行归纳、整理。  相似文献   

9.
目的评估乙型肝炎病毒血清标志物(HBV—M)胶体金免疫层析法(GICA)的分析性能。方法收集雅培i1000免疫检测系统检测的HBV—M阳性标本653例、阴性标本103例和HBV—M标准品,以化学发光免疫测定法(CLIA)检测结果为标准,分别用酶联免疫吸附试验(ELISA)试剂和5种GICA试剂检测,统计分析检测结果。结果检测HBV—M敏感性:ELISA和GICA检测HBV—M的最低检测限分别为乙型肝炎表面抗原(HBsAg)0.5IU/mL和4~8IU/mL、乙型肝炎表面抗体(抗HBs)10mlU/mL和15~30mlU/mL、乙型肝炎e抗原(HBeAg)1NCU/mL和2~4NCU/mL、乙型肝炎e抗体(抗HBe)2NCU/mL和64~256NCU/mL、乙型肝炎核心抗体(抗HBc)2IU/mL.和32—256IU/mL。检测标本HBV—M的符合率:ELISA分别为87.4%、99.0%、87.7%、96.6%、100.0%,GICA分别为69.6%~71.7%、69.0%~72.0%、70.5%~73.9%、42.2%~49.1%、50.0%~60.0%。检测HBV—M特异性:除GICA柃测抗HBs的特异性略差外,其余各HBV—M的检测特异性均是可接受。,结论GICA检测HBV—M低浓度的标本,漏检率很高,只能用于GICA检测线性范围里HBsAg的筛查,不能作为临床诊断乙型肝炎和疗效考核的依据。  相似文献   

10.
目的比较化学发光免疫分析(CLIA)和酶联免疫吸附试验(ELISA)检测血清HBsAg及HBsAb的结果。方法采用CLIA法和ELISA法测定定值血清中的HBsAg和HBsAb,并对两种结果进行比较和分析。结果测定HBsAg时,ELISA法的批内CV:1IU/ml为9.6%,2IU/ml为7.3%,批间CV:1IU/ml为17.8%,2IU/ml为15.1%,灵敏度为0.25IU/ml,CLIA法的批内CV:1IU/ml为2.33%,2IU/ml为1.07%,批间CV:1IU/ml为6.57%,2IU/ml为3.78%,灵敏度为0.063IU/ml。HBsAb测定时,ELISA法的批内CV:10mIU/ml为11.8%,20mIU/ml为8.7%,批间CV:10mIU/ml为18.6%,20mIU/ml为15.9%,灵敏度为10mIU/ml,CLIA法的批内CV:10mIU/ml为4.79%,20mIU/ml为3.52%,批间CV:10mIU/ml为7.58%,20mIU/ml为6.13%,灵敏度为2.5mIU/ml。结论 ELISA法敏感度相对低,精密度差;CLIA法灵敏度高,精密度好,且自动化程度高,检测更快速,更适合于临床。  相似文献   

11.
目的 对西门子Centaur cp全自动化学发光分析仪测定超敏促甲状腺素(TSH3-Ultra)分析灵敏度的性能进行验证.方法 用促甲状腺素配套稀释液作为空白样品,并以此稀释液对低浓度(浓度值为0.010 0 mIU/L)的新鲜患者血清进行稀释,配置成低浓度系列样品作为分析灵敏度的实验样品.空白样品批内重复测定10次,其他系列样品日间重复测定10次,记录每次检测的发光强度(RLUs),计算检测低限、生物检测限、功能灵敏度等.结果 按99.7%可信限计算,TSH3-Ultra检测低限为0.001 0 mIU/L;按照国际纯粹和化学联合会的规定TSH3-Ultra检测低限为0.001 1 mIU/L,与生产厂家产品说明书声明一致;生物检测限为0.005~0.006 mIU/L,功能灵敏度为0.006 mIU/L,低于生产厂家产品说明书声明的功能灵敏度(0.008 mIU/L).结论 西门子Centaur cp全自动化学发光分析仪检测TSH3-Ultra分析灵敏度的性能与厂家声明性能基本一致.  相似文献   

12.
BACKGROUND: According to recent International Organization for Standardization (ISO) standards, the limit of detection (LoD) of an assay should be estimated taking both type I (alpha) and II (beta) errors into account. The suggested procedure, however, supposes gaussian distributions of both blank and sample measurements and a linear calibration curve. In clinical chemistry, asymmetric, nongaussian blank distributions are common, and the calibration curve may be nonlinear. We present a partly nonparametric procedure that takes these aspects into account. METHODS: Using theoretical distribution models and simulation studies, we developed a LoD estimation procedure suitable for the field of clinical chemistry that is partly based on nonparametric statistics. RESULTS: For sample size n, the nonparametrically determined 95th percentile of the blank measurements obtained as the value of the [n(95/100) + 0.5]th ordered observation defines the limit for results significantly exceeding zero [limit of blank (LoB)]. The LoD is the lowest value that is likely to yield a result exceeding the LoB. LoD is estimated as: LoB + cbeta x SDS, where SDS is the analytical SD of a sample with a low concentration; cbeta = z(1 - beta)/[1 - 1/(4 x f)]; z(1 - beta) is the standard normal deviate; and f is the number of degrees of freedom for estimation of SD(S). c(beta) is approximately equal to 1.65 for a type II error of 5%. Approaches and needed tabular values for calculation of confidence limits are presented as well as sample size. Worked examples are given to illustrate estimation and verification of the limit of detection. Simulation results are used to document performance. CONCLUSION: The proposed procedure appears useful for application in the field of clinical chemistry and promotes a standardized approach for estimating LoDs of clinical chemistry assays.  相似文献   

13.
目的:试用实时荧光定量 PCR 定量检测乙型肝炎病毒核酸(HBV DNA),求出其线性范围与最低检出限,从而对本实验室检测 HBV DNA 的主要性能进行验证。方法参照相关的文件,通过系列稀释高浓度标本得到超过厂家申明线性范围的系列浓度标本,进行线性范围的验证;通过系列稀释低浓度标本得到超过厂家申明最低检出限的系列浓度标本,进行最低检出限的验证。结果 HBV DNA 检测的线性范围为8.58&#215;102~8.41&#215;107 IU/mL,最低检出限为4.07&#215;102 IU/mL。结论实时荧光定量 PCR 检测 HBV DNA 的线性范围与最低检出限达到预期要求,检测方法和验证方案简便、可行。  相似文献   

14.
目的 探讨西门子ADVIA Centaur 240化学发光免疫系统检测B型利钠肽(BNP)的检测限值和功能灵敏度.方法 参照美国临床实验室标准化协会(CLSI)EP-17A文件和有关文献,选择BNP通用稀释液作为空白标本,批内重复测定10次,计算光量子值的均数、标准差和变异系数(CV),用以确定检测低限(LLD);同时,空白标本每天上、下午各检测1批,每批重复检测3次,连续检测5 d,共获得30个空白结果,用非参数方法确定空白限(LoB).制备接近1~4倍LoB浓度的系列实验标本,各浓度每天上、下午各检测1次,连续检测5 d,计算每个浓度水平光量子值的均数、标准差和CV,用以确定生物检测限(BLD)和CV值为10%条件下的功能灵敏度(FS);同时对所有80个测定结果,按EP-17A文件采用非参数方法确定检出限(LoD).结果 ADVIA Centaur 240检测系统BNP的LLD为1.47 pg/mL,LoB为0.86 pg/mL,均低于厂家声明的灵敏度;BLD为3.27~3.71 pg/mL,LoD为3.54 pg/mL,10%CV条件下的FS为4.64 pg/mL.结论厂商声明的LLD和BLD得到验证,同时建立了实验室BNP检测的LoB、LoD和FS,为临床提供了准确、可靠的检验结果.建议临床实验室使用EP-17A文件确定LoB和LoD,并确定在一定精密度条件下的FS.  相似文献   

15.
目的 评价电化学发光免疫法检测NT-proBNP的分析性能.方法 参考CLSI发布的方法学评价系列文件,对Roche Cobas E601电化学发光免疫系统检测NT-proBNP的精密度、正确度、空白限(LoB)和检出限(LoD)、功能灵敏度(FS)、分析测量范围(AMR)、最大稀释度、临床可报告范围(CRR)以及胆红素、血红蛋白和乳靡对分析的干扰进行评价,并与厂商声明的性能和有关的质量标准进行比较.根据CLSI C28-A2文件,选择18~74岁的健康体检者80名,按年龄分为4组,每组20名,以验证厂商声明的生物参考区间的适用性.结果 电化学发光免疫法检测NT-proBNP的批内CV和总CV分别为1.1% ~2.2%和1.5% ~2.9%,与卫生部临床检验中心室间质量评价调查品同方法组的均值和配套校准品标示值的偏差分别为2.7% ~5.9%和2.7%~7.5%,LoB、LoD和FS分别为2.5、7.8和8.8 pg/ml,AMR为8~35 126 pg/ml,最大允许稀释度为1∶2,CRR为9~70 252 pg/ml,胆红素(428 μmol/L)、血红蛋白(2 g/L)和乳靡(2 200 FIU)对NT-proBNP测定无干扰.不同年龄组各20名健康参考个体NT-proBNP浓度除1个检测值(167 pg/ml)超出厂商提供的参考区间外,其余均在厂商提供的参考区间内.结论Roche Cobas E601电化学发光免疫系统检测NT-proBNP的主要分析性能达到了厂商声明的性能和有关的质量要求,建立的LoD、FS、最大稀释度和CRR可为临床提供更好的质量保证,厂商声明的生物参考区间可满足临床需要.  相似文献   

16.
The absorption of erythropoietin (EPO) from rat small intestine was studied using gastro-intestinal patches (GI-PS) in the presence of absorption enhancers. Surfactants such as a saturated polyglycolysed C8-C18 glyceride (Gelucire 44/14), PEG-8 capryl/caprylic acid glycerides (Labrasol), and polyoxyethylene hydrogenated castor oil derivative (HCO-60) were used as absorption enhancers at 143, 94 and 20 mg/kg, respectively. The absorption of EPO was studied by measuring serum EPO levels by an ELISA method after small intestinal administration of EPO-GI-PS preparation in rats at the EPO dose level of 100 IU/kg. Labrasol showed the highest absorption enhancing effect after intrajejunum administration with maximum serum EPO level of 84.1+/-11.4 mIU/ml while Gelucire 44/14 and HCO-60 showed 43.5+/-9.8 and 26.5+/-2.3 mIU/ml, respectively. The appropriate site for EPO absorption was also investigated. Jejunum was found to be the most efficient absorption site for the absorption of EPO from GI-PS. Using Labrasol as the absorption enhancer and jejunum as the absorption site, the effect of EPO dose on EPO absorption was studied by increasing the EPO dose from 50, to 100, 300 and 600 IU/kg. It was found that 100 IU/kg was the optimum dose with a serum EPO level of 84.1+/-11.4 mIU/ml while escalating doses showed decreases in serum EPO levels 48.3+/-5.6 for 300 IU/kg and 50.6+/-10.3 mIU/ml for 600 IU/kg. The percent bioavailability (BA) of EPO-GI-PS with Labrasol as absorption enhancer was 7.9 at 50 IU/kg, 12.1 at 100 IU/kg, 3.2 at 300 IU/kg and 1.2 at 600 IU/kg. Histological studies showed no adverse effect at the site of administration.  相似文献   