1.
目的:比较Goldmann压平眼压计(Goldmannapplanationtonometer,GAT)与非接触眼压计(non-contacttonometer,NCT)测量眼压的差异,以评价NCT与GAT测量的相关性。方法:对265例志愿者(529眼)分别采用Goldmann压平眼压计与非接触眼压计测量眼压。结果:非接触眼压计的测量结果低于Goldmann压平眼压计,且差异有显著性(19.13vs23.43,t=22.644,P<0.01),随眼压值的升高,两者相差幅度增大,差异在眼压〉30mmHg时更为明显,但相关系数逐渐变小。结论:非接触眼压计眼压测量值较Goldmann眼压测量值偏低,非接触眼压计眼压值为临界眼压时,需应用Gold-mann压平眼压计校正,以便及时发现病理性眼压升高,避免青光眼的漏诊和失治。  相似文献   

2.
目的::对比分析 Goldmann 压平式眼压计( Goldmann applanation tonometer, GAT)和非接触眼压计( non-contact tonometer,NCT)测量青光眼患者眼压的结果,寻找其规律性。方法:分别使用GAT和NCT测量108例206眼青光眼患者的眼压。结果:青光眼患者108例206眼的平均眼压:GAT测量为29.77±10.27mmHg,NCT测量为24.59±8.58mmHg,两者比较存在显著性差异(P<0.01),而且眼压越高,差异性越显著。结论:NCT 测量值较 GAT 测量值低,眼压越高,差值越大。建议使用GAT检查青光眼患者的眼压,以免漏诊和贻误治疗。  相似文献   

3.
非接触式眼压计与Goldmann压平眼压计测量眼压的比较   总被引:8,自引:2,他引:6  
目的 :比较非接触式眼压计 ( NCT)和 Goldm ann压平眼压计测量眼压的差异。方法 :对 112例 ( 2 2 0眼 )志愿者分别进行 NCT和 Goldmann压平眼压计眼压测量及中央角膜厚度测量 ,并对其中 16 8眼进行自动验光检查。结果 :NCT和 Goldmann眼压计眼压测量值分别为 1.98± 0 .6 9k Pa和 2 .34± 0 .77k Pa。 NCT眼压测量值较 Goldmann眼压测量值偏低 0 .36± 0 .37k Pa( P<0 .0 5 )。眼压在 1.33~ 2 .6 7k Pa内 ,二种眼压计测量眼压值偏差最小。角膜厚度和眼球屈光度与 NCT眼压测量值分别呈明显正相关和负相关。结论 :NCT眼压测量值较Goldmann眼压测量值偏低 ,NCT眼压值为临界眼压时 ,应应用 Goldm ann压平眼压计校正  相似文献   

4.
非接触眼压计测量的评价——与Goldmann眼压计的比较   总被引:11,自引:2,他引:9  
目的 :对比非接触式眼压计 (NCT)与Goldmann眼压计所测量的眼压值以评价非接触式眼压计在临床的应用价值。方法 :10 5例 2 0 9只眼纳入本观察 ,其中 14 2只眼作了验光、眼轴及角膜曲率测量 ,并调查患者对两种眼压计的接受度。结果 :两种眼压计测量结果的相关系数为 0 975 ,两者差异有显著性 (t值 -7 949,P <0 0 0 1)。非接触式眼压计的测量结果高于Goldmann眼压计 ,尤其在眼压 >2 1mmHg组更为明显 (t值 -5 5 0 6,P <0 0 0 1)。角膜曲率、屈光度、散光、眼轴与两者测量值间差d无相关性 ,相关系数分别为 -0 0 5 4,-0 0 41,-0 13 5 ,0 113 ,P值均 >0 0 5。患者对非接触式眼压计的接受程度大于Goldmann眼压计结论 :非接触式眼压计可用于临床普查 ,但在青光眼的临床工作中建议使用Goldmann压平式眼压计  相似文献   

5.
手持式非接触眼压计测量值准确性的比较   总被引:1,自引:0,他引:1  
目的:对手持式非接触眼压计测量眼压的准确性及可靠性进行评价。方法:分别采用keeler,手持非接触眼压计和R900,Goldmann压平眼压计对58例患者(113只眼)的限内压进行了测量并对两种测量方法的测量结果进行了比较。结果:采用手持非接触眼压计测得的平均眼压为16.31±5.59mmHg(1mmHg=0.1333kpa),采用Goldmann压平眼压计测得的平均眼压值为17.49±6.13mmHg。经统计学t检验,两种测量结果差别无统计学意义(P>0.05)。经直线相关和回归分析发现两种方法测量值间存在正相关(r=0.8942,b=0.8154)。结论:手持非接触眼压计同Goldmann压平眼压计相比具有同样的精确性和可靠性,可用于青光眼临床和青光眼筛选。眼科学报 1995;11:86—88。  相似文献   

6.
角膜厚度对两种眼压测量方法的影响   总被引:6,自引:2,他引:4  
目的 比较非接触眼压计 (non -contacttonometer ,NCT)和Goldmann压平眼压计测量眼压的差异 ,并分别探讨中央角膜厚度 (centralcornealthickniss ,CCT)对这两种测量方法的影响。方法 对 1 0 8例拟接受PRK或LASIK手术的患者行CCT ,NCT和Goldmann压平计眼压测量。结果NCT和Goldmann压平眼压计测得的眼压均值具有显著性差异 (F =89 .70 4 4,P <0 . 0 1 )。CCT与NCT和Goldmann压平眼压计测量值呈正相关 ,相关系数分别是r =0. 4 96 0 (t =8 .356 3,P <0 .0 0 1 )和r =0 . 2 1 1 3(t =3. 1 6 2 3,P <0 .0 0 1 )。结论 NCT和Goldmann压平眼压计测量眼压值有差异 ,NCT测量值大于Goldmann压平眼压计 ,CCT对NCT的影响大于Goldmann压平眼压计。  相似文献   

7.
目的了解动态轮廓眼压计的临床实用性。方法67例患者134眼随即先后用动态轮廓眼压计与Goldmann压平眼压计进行眼压测量,比较两种眼压计测量值的差异。结果134眼中,用Goldmann压平眼压计测量的眼压平均值为(15.82±-4.88)mmHg,用动态轮廓眼压计测量的眼压平均值为(14.5±-4.33)mmHg,;两者无明显差异。但73%的患者的动态轮廓眼压计的眼压值高于Goldmann压平眼压计的测量值。两者的相关性为0.83。结论动态轮廓眼压计可以较好的反映患者的眼压,操作简便,值得推广应用。  相似文献   

8.
目的比较我们自行研制的凸面眼压计与Goldmann压平眼压计(以下简称Goldmann眼压计)所测量的C值,以评价凸面眼压计在临床的应用价值。方法分别应用凸面眼压计与Goldmann眼压计测量15例(30只眼)正常人的C值;15例(15只眼)青光眼患者的C值,并作统计学分析。结果凸面眼压计C值是Goldmann眼压计C值的两倍,两种眼压计C值测量结果差异有显著性意义(P<0.01)。结论用凸面眼压计所测得的C值更接近于流体压力方法测得的C值,因此凸面眼压计更适合于眼压描记,具有临床使用价值。  相似文献   

9.
目的为获得眼压测量的更大精确性,设计并研制一架新型凸面眼压计。方法根据Goldmann压平眼压计的基本原理,将Goldmann压平眼压计测压头的平面改成曲率半径为7.8mm的凸面,并在凸面中央刻一个3.06mm直径的圆形刻线,作为压迫角膜固定面积的标志。其测量方法与Goldmann压平眼压计相同。用本眼压计对55例(109只眼)进行了初步眼压测试,并与Goldmann压平眼压计进行对照。结果凸面眼压计的平均眼压为(16.08±5.13)mmHg,Goldmann压平眼压计平均眼压为(14.96±4.27)mmHg,t值=0.28,P>0.05,两者差异无显著统计学意义;r=0.82,两者有较好的正相关。结论本眼压计有较高的准确性,可用作临床眼压测量。  相似文献   

10.
程玲艳  崔娟莲  段宣初 《眼科》2011,20(1):33-37
目的探讨动态轮廓眼压计(DCT)与Goldmann压平眼压计(GAT)及非接触眼压计(NCT)测量眼压的准确性,并比较三种眼压计测量结果与中央角膜厚度(CCT)的相关性。设计前瞻性、比较性病例系列。研究对象连续选取90例(90眼)10~76岁正常人。方法采用KONAN非接触式角膜内皮镜测量CCT后,对所有入选者单眼以随机顺序采用Pascal型DCT、GAT及Topcon型NCT测量眼压。测量结果两两比较,并将眼压值与CCT进行直线回归分析。主要指标眼压值,Pearson相关系数。结果 90例正常人DCT眼压平均值(17.33±2.71 mm Hg)明显高于GAT(14.27±2.81 mm Hg)(P=0.000)及NCT(14.67±2.93 mm Hg)(P=0.000),平均差异分别为(3.06±2.01)mm Hg和(2.67±2.20)mm Hg;GAT与NCT之间平均差异为(-0.39±2.29)mm Hg(P=0.105)。DCT与GAT眼压值之间相关系数r=0.736(P=0.000);与NCT眼压值之间相关系数r=0.699(P=0.000)。GAT、NCT眼压值与CCT均明显相关(r=0.370,P=0.000;r=0.508,P=0.000);DCT眼压值与CCT无明显相关性(r=0.051,P=0.639)。DCT和GAT的差值与年龄无明显相关性(r=0.064,P=0.052)。结论 DCT测量的眼压值虽高于GAT及NCT,但不受CCT的影响,可能较GAT和NCT测量的眼压值更接近真实值。  相似文献   

11.

目的:分析非接触式眼压计(non contact tonometer,NCT)、ICare回弹式眼压计(ICare rebound tonometer,RBT)在青光眼患者中眼压测量结果的准确性。

方法:回顾性病例对照研究。选取青光眼患者113例185眼。用RBT、NCT和Goldmann眼压计(GAT)分别于9:00、16:00两次进行眼压测量并记录为两组数据,分别用秩和检验、国际标准8621指南评估其准确性,Bland-Altman一致性分析两组测量结果的一致性,并以GAT测量值作为标准将数据分为异常眼压组(<10mmHg或>21mmHg)和正常眼压组(10~21mmHg),在不同组间分析其一致性。应用Spearman相关分析眼压计间的相关性。

结果:GAT、NCT和RBT三组测量值间存在差异(P<0.01); GAT获得的眼压测量值与RBT和NCT测量值间存在相关性(rs=0.71、0.77,P<0.001)。NCT与GAT测量值接近(P=0.92),而RBT与GAT相比,眼压测量值较高(P<0.05)。然而NCT、RBT同GAT相比一致性界限范围均较宽,分别为-6.2~6.0、-5.2~7.6mmHg,在正常眼压组一致性界限范围分别为-5.9~5.9、-4.3~7.5mmHg,在非正常眼压组分别为-7.3~6.4、-7.5~5.6mmHg。根据ISO 8612指南进行评价,NCT和RBT在三组中超过95%的一致性限制的异常值分别为3.9%、11.3%、12.2%和26.3%、11.3%、12.2%。

结论:NCT和RBT均不能简单代替GAT用于青光眼患者眼压的测量。在青光眼患者中,随着测量值偏离正常眼压范围,NCT和RBT的测量误差也有所增大。  相似文献   


12.
PURPOSE: To determine the agreement between the measurement of intraocular pressure (IOP) by the rebound tonometer (RBT) and by the Goldmann applanation tonometer (GAT) and to find out the effect of central corneal thickness (CCT) values on IOP measurements in glaucoma patients. METHODS: IOP was measured with the RBT and GAT, respectively, in 61 eyes of 61 glaucoma patients. CCT was measured using an ultrasonic pachymeter after all IOP determinations had been made. The mean IOP measurement by the RBT was compared with the measurement by the GAT, by Student's t-test. Bland-Altman analysis was performed to assess the clinical agreement between the two methods. The effect of CCT on measured IOP was explored by linear regression analysis. RESULTS: The mean patient age was 56.7+/-21.1 years (range: 30-80 years). There were 32 (52.46%) women and 29 (47.54%) men in the study group. The mean IOP readings were 18.70+/-4.76 mmHg using the RBT, and 18.27+/-3.49 mmHg using the GAT. The difference was not statistically significant (mean difference 0.43+/-2.55, P=0.2). A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <2.3 mmHg between the RBT and GAT. There was a strong correlation between the RBT and GAT readings (r=0.852, P<0.0001). The IOP measurements with the two methods were correlated with CCT (r=0.40, P=0.02 for the RBT and r=0.48, P<0.0001 for the GAT). The IOP increased 1.1 mmHg and 8 mmHg for every 100-microm increase in CCT for the GAT and RBT, respectively. CONCLUSION: The RBT slightly overestimated the IOP value by 0.43 mmHg on average when compared with the GAT. Nevertheless, the RBT readings appeared to be more affected by the various thicknesses of different corneas when compared with those obtained using the GAT.  相似文献   

13.
目的:验证动态轮廓眼压计(DCT)的临床性能。方法:对81例143眼青光眼及可疑病例青光眼患者用DCT测量眼内压(IOP);132眼同时用Goldmann压平眼压计(GAT)测量眼压,部分病例同时用NIDEKUP-1000型角膜测厚仪测量中央角膜厚度(CCT)。GAT与DCT测量结果采用配对t检验,GAT、DCT测量值与CCT的关系及DCT测量值与眼脉动振幅(OPA)的相关关系采用Spearman双变量相关分析。结果:①DCT和GAT测得的眼压均数分别为(18.6±3.9)mmHg和(17.4±4.1)mmHg;DCT测得的眼压值高于GAT,其差值均数为(1.1±2.3)mmHg;两种眼压计测得的眼压值呈正相关(r=0.83,P<0.01)。②116眼同时完成了DCT眼压测量及CCT测量,两者相关系数r=0.03,P=0.77;113眼同时完成了GAT眼压测量及CCT测量,两者相关系数r=0.28,P=0.003。③143眼同时记录了DCT眼压值和眼脉动振幅(OPA),其均数分别为(18.6±3.9)mmHg和(2.6±1.1)mmHg,两者的相关系数r=0.32,P<0.01。结论:以上结果初步证实:①DCT眼压计测量值与GAT眼压测量值高度相关,但DCT测量值略高于GAT测量值,提示DCT可用于临床诊断。②DCT测量值与CCT不相关,GAT测量值与CCT显著相关,提示DCT在青光眼诊断中有独特优势。③DCT测量的OPA与IOP值显著相关,相关的机制及临床意义有待探讨。  相似文献   

14.

Background

To compare intraocular pressure (IOP) measurements obtained with the Topcon CT‐80 non‐contact tonometer (NCT) and Goldmann applanation tonometer (GAT), in different ranges of IOP in normal and glaucoma subjects, and to assess the influence of central corneal thickness (CCT) on the IOP measurements in Asian Indian eyes.

Methods

Four hundred and two eyes of 402 subjects (193 newly diagnosed primary open angle glaucoma [POAG] and 209 normal) were enrolled for this prospective study. For each eye, IOP was measured with GAT by a glaucoma specialist and NCT by a trained optometrist. The IOP values were compared among the tonometers in the three different IOP ranges (≤ 12 mmHg, 13–20 mmHg and ≥ 21 mmHg) using Bland–Altman graphs. Correlation between GAT and NCT was assessed by Pearson correlation co‐efficient. CCT was measured with ultrasound pachymetry and its correlation with GAT and NCT was analysed using linear regression analysis.

Results

The mean paired difference of IOP between NCT and GAT was 1.556 ± 2.69 mmHg (r = 0.26, p = 0.006) at IOP range of ≤ 12 mmHg, ?1.665 ± 2.6 mmHg (r = 0.51, p < 0.0001) in IOP range of 13–20 mmHg and ?2.202 ± 3.44 mmHg (r = 0.82, p < 0.0001) in the IOP range of ≥ 21 mmHg. Linear regression analysis showed a mean IOP variation of 0.27 mmHg per 10 μm change in CCT for NCT (p < 0.0001) and IOP change of 0.19 mmHg per 10 μm change in CCT for GAT (p = 0.01).

Conclusion

In this study of normotensive and POAG subjects, the Topcon CT‐80 NCT showed an overestimation of IOP at the lower range and underestimation of IOP in normal and higher ranges of IOP. Clinicians should keep in mind that CCT influences IOP measurement with both types of tonometer and that the IOP readings obtained with these tonometers are not interchangeable.
  相似文献   

15.
目的:比较不同注视性质非接触式眼压计(NCT)和Goldmann眼压计(GAT)测量眼压结果的差异,以评价NCT对偏心注视患者眼压测量结果的准确性.方法:用NCT和GAT分别测量54例108眼中心注视患者及49例61眼偏心注视患者的眼压,并将各测量结果进行统计学处理.结果:中心注视组54例108眼GAT测量平均为21.30mmHg,NCT测量平均为21.21mmHg,经统计学处理差异无显著性;偏心注视49例61眼GAT测量平均为21.10mmHg,NCT测量平均为25.33mmHg,经统计学处理,差异具有显著性(P<0.05),两种眼压计的差值与GAT测量值成正相关,相关系数是0.792.结论:NCT的测量值高估了偏心注视患者的眼压,检查注视性质有助于正确的评估NCT所测得的眼压结果,临床工作中建议使用GAT对偏心注视患者的NCT眼压值进行矫正.  相似文献   

16.
目的 比较非接触眼压计(气流眼压)、Goldmann压平眼压计(压平眼压)、Icare pro回弹眼压计(回弹眼压)、Tono-pen AVIA眼压计(笔式眼压)和可视化角膜生物力学分析仪(生物力学眼压)测量眼压的一致性,进一步分析眼压和角膜生物力学的相关性。设计 诊断方法评价。研究对象 健康志愿者44例。方法 同一医师每天同一时段分别应用上述五种眼压计测量受试者的左眼眼压。将眼压与中央角膜厚度、角膜第一压平时间进行Pearson相关分析。主要指标 眼压值、眼压差值的平均值、相关系数。结果 压平眼压、气流眼压、回弹眼压、笔式眼压和生物力学眼压5种眼压结果分别为(15.9±3.3)mmHg、(14.8±2.9)mmHg、(16.9±3.3)mmHg、(14.7±2.5)mmHg和(16.1±3.0)mmHg。其中,气流眼压、笔式眼压较压平眼压低(P=0.01,0.00)。气流眼压、回弹眼压、笔式眼压、生物力学眼压均与压平眼压正相关(r=0.63、0.37、0.63、0.55,P均<0.05);Bland-Altman分析两种测量方式眼压差值平均值分别为:气流眼压与压平眼压:-1.1 mmHg [95%一致性界限(95% LoA)为-6.4,4.2],笔式眼压与压平眼压:-1.2 mmHg(95% LoA -6.3,3.9),回弹眼压与压平眼压:1.0 mmHg(95% LoA为-6.2,8.3),生物力学眼压与压平眼压:0.3 mmHg(95% LoA为-5.6,6.2)。所有研究对象中央角膜厚度(550.5±29.2)μm,角膜第一压平时间(7.63±0.36)ms。笔式眼压和中央角膜厚度呈正相关(r=0.40,P=0.01)。五种测试的眼压均与角膜第一压平时间呈正相关(r=0.53,0.64,0.55,0.46,0.99;P均<0.05)。结论 Icare pro和Corvis ST测量眼压与Goldmann眼压计无明显差异,气流眼压、笔式眼压较压平眼压略低,Corvis ST与Goldmann眼压一致性最好。角膜第一压平时间是影响眼压测量结果的重要因素,时间越长,眼压测量值越高。(眼科, 2020, 29: 365-369)  相似文献   

17.
目的:比较采用Goldmann压平眼压计(Goldmann applanation tonometer,GAT)、非接触眼压计(non-contact tonometer,NCT)和Schiotz眼压计(Schiotz tonometer,ST)的眼压(intraoeular pressure,IOP)测量,评估角膜中央厚度(central corneal thickness,CCT)对读数的影响。方法:使用GAT、NCT和ST对所有患者的右眼进行眼压测量。超声角膜厚度测量法测定CCT。所有IOP及CCT测量由同一检查者进行。计算CCT25%(Q1)百分位数和75%(Q3)百分位数值,并通过这种方法将该组分为薄、中、厚角膜亚组。使用Statplus软件进行统计分析。结果:全系列144眼,GAT测量平均IOP为17.4±4.9mmHg,NCT为16.0±5.8mmHg,ST为14.0±4.0mmHg(Friedman方差分析P<0.01)。IOP水平和CCT之间的相关系数NCT为0.787(P<0.01),GAT为0.630(P<0.01),ST为0.565(P<0.01)。ST测量中,纠正的IOP误差和CCT之间的相关性在厚角膜明显弱(r=0.381,P=0.022)。结论:NCT是最易受不同CCT影响的设备。ST读数似乎比GAT和NCT读数受CCT的影响小。特别是在厚角膜,与NCT和GAT相比,ST可以被认为是一个更可靠的仪器。  相似文献   

18.
目的:在原发性开角型青光眼( POAG )中应用佳能TX-F非接触式眼压计( NCT)和Goldmann压平眼压计( GAT)测量眼压( IOP),并比较测量值。 方法:55例(55右眼)确诊为POAG的患者接受详细的眼科检查,光学相干断层扫描成像和自动视野检查。使用NCT1(一次喷气模式), NCT3(三次喷气模式)和GAT测量眼压,每隔5分钟一次。 结果:55例(55右眼) POAG 患者平均年龄为64.1±8.1岁。比较NTC1,NTC3测量的眼压值(14.22±3.42,14.28±3.29mmHg)与GAT测量的眼压值(14.66±3.49mmHg)无统计学差异( P=0.291)。使用Bland-Altman 方法比较NCT1-GAT, NCT3-GAT和NCT1-NCT3得出的95%一致性界限(IOA)分别为-4.9~+4.4mmHg,-4.1~+3.4mmHg和-3.4~+3.3 mmHg。 结论:虽然NCT与GAT测量的眼压值相似,但偏大的IOA范围限制了NCT1,NCT3和GAT在POAG患者中的互换应用。  相似文献   