光学表面成像系统实时运动监测精度研究
引用本文:陈黎1,2,白龙1,李光俊1,全红2,柏森1. 光学表面成像系统实时运动监测精度研究[J]. 中国医学物理学杂志, 2021, 0(9): 1053-1056. DOI: 10.3969/j.issn.1005-202X.2021.09.001
作者姓名:陈黎1  2  白龙1  李光俊1  全红2  柏森1
作者单位:1. 四川大学华西医院肿瘤中心放疗科,四川成都610041;2. 武汉大学物理科学与技术学院,湖北武汉430072
摘    要:目的:探究光学表面成像系统实时运动监测的精度。方法:将30例患者的呼吸曲线输入到模体中模拟呼吸运动,同时利用Catalyst系统对模体进行实时运动监测,比较系统监测的呼吸曲线与参考曲线,从而得到光学表面成像系统实时运动监测的精度。结果:光学表面成像系统监测的呼吸曲线与参考曲线具有较高的一致性,相关系数均大于0.99,显著相关。监测误差的平均值为(0.24±0.04)mm,并且随着呼吸信号频率的增加而减小。结论:光学表面成像系统的实时运动监测精度较高,可用于对患者呼吸运动的监测。在进行呼吸门控治疗时,应考虑呼吸监测系统引入的误差。

关 键 词:光学表面成像系统  呼吸运动  监测精度

Real-time motion tracking accuracy of optical surface imaging system
CHEN Li1,2,BAI Long1,LI Guangjun1,QUAN Hong2,BAI Sen1. Real-time motion tracking accuracy of optical surface imaging system[J]. Chinese Journal of Medical Physics, 2021, 0(9): 1053-1056. DOI: 10.3969/j.issn.1005-202X.2021.09.001
Authors:CHEN Li1  2  BAI Long1  LI Guangjun1  QUAN Hong2  BAI Sen1
Affiliation:1. Department of Radiation Oncology, Cancer Center, West China Hospital, Sichuan University, Chengdu 610041, China 2. School ofPhysics and Technology, Wuhan University, Wuhan 430072, China
Abstract:Objective To explore the accuracy of real-time motion tracking using optical surface imaging system. MethodsThe respiratory curves of 30 patients were input into a phantom for simulating respiratory motion, and Catalyst system wasused for tracking the real-time motion of the phantom. The respiratory curves recorded by the system were compared with thereference curves, thereby obtaining the real-time motion tracking accuracy of optical surface imaging system. Results Therespiratory curves recorded by the system had a high consistency with the reference curves, and the correlation coefficientwas higher than 0.99, indicating significant correlation. The average of tracking errors was (0.24±0.04) mm, and the error wasdecreased while the respiratory signal frequency was increased. Conclusion The optical surface imaging system which has arelatively high real-time respiratory motion tracking accuracy can be used for respiratory motion tracking. During respiratorygating treatment, the error introduced by the respiratory motion tracking system should be considered.
Keywords:optical surface imaging system respiratory motion tracking accuracy