Mirau-based line-field confocal optical coherence tomography for three-dimensional high-resolution skin imaging |
| |
Authors: | Weikai Xue Jonas Ogien Pavel Bulkin Anne-Lise Coutrot Arnaud Dubois |
| |
Affiliation: | aUniversité Paris-Saclay, Institut d’Optique Graduate School, CNRS, Laboratoire Charles Fabry, Palaiseau, France;bDAMAE Medical, Paris, France;cLPICM, CNRS, Ecole Polytechnique, Institut Polytechnique de Paris, Palaiseau, France |
| |
Abstract: | SignificanceLine-field confocal optical coherence tomography (LC-OCT) is a recently introduced high-resolution imaging modality based on a combination of low-coherence optical interferometry and reflectance confocal optical microscopy with line illumination and line detection. Capable of producing three-dimensional (3D) images of the skin with cellular resolution, in vivo, LC-OCT has been mainly applied in dermatology and dermo-cosmetology. The LC-OCT devices capable of acquiring 3D images reported so far are based on a Linnik interferometer using two identical microscope objectives. In this configuration, LC-OCT cannot be designed to be a very compact and light device, and the image acquisition speed is limited.AimThe objective of this work was to develop a more compact and lighter LC-OCT device that is capable of acquiring images faster without significant degradation of the resolution and with optimized detection sensitivity.ApproachWe developed an LC-OCT device based on a Mirau interferometer using a single objective. Dynamic adjustment of the camera frequency during the depth scan is implemented, using a faster camera and a more powerful light source. The reflectivity of the beam-splitter in the Mirau interferometer was optimized to maximize the detection sensitivity. A galvanometer scanner was incorporated into the device for scanning the illumination line laterally. A stack of adjacent B-scans, constituting a 3D image, can thus be acquired.ResultsThe device is able to acquire and display B-scans at 17 fps. 3D images with a quasi-isotropic resolution of (1.3, 1.9, and in the , and directions, respectively) over a field of () can be obtained. 3D imaging of human skin at cellular resolution, in vivo, is reported.ConclusionsThe acquisition rate of the B-scans, at 17 fps, is unprecedented in LC-OCT. Compared with the conventional LC-OCT devices based on a Linnik interferometer, the reported Mirau-based LC-OCT device can acquire B-scans times faster. With potential advantages in terms of compactness and weight, a Mirau-based device could easily be integrated into a smaller and lighter handheld probe for use by dermatologists in their daily medical practice. |
| |
Keywords: | optical coherence tomography optical imaging confocal microscopy dermatology skin cancer |
|
|