Early results of a new mechanical tri-leaflet heart valve prosthesis - Tricusp: an animal study |
| |
Authors: | Kenneth Pehrsson S. Maria J. Eriksson Jan C. Lentell Staffan Bowald |
| |
Affiliation: | 1. Department of Cardiology Karolinska Hospital Stockholm Sweden;2. Department of Thoracic Radiology and Physiology Karolinska Hospital Stockholm Sweden;3. JCL Technic Ltd Stockholm;4. Assist Medical AB Hagby Sweden |
| |
Abstract: | Objective --A new tri-leaflet mechanical heart valve made of titanium was inserted in the mitral position to evaluate early results. Methods and results --Five sheep were followed between 3 and 6 months (mean 4.7 months) and performed very well clinically during the follow-up period. In three of five animals a minor para-valvular leakage without hemodynamic importance was observed. The invasive pressure measurements at the end of follow-up revealed only minor transvalvular pressure differences between 1 and 4 r mmHg (mean 2.4 r mmHg). Similar results were obtained by echo-Doppler technique. Histological examination of the tissue around the valve showed no signs of foreign body reaction or chronic inflammatory reaction. Conclusion --This new tri-leaflet heart valve has shown an excellent hemodynamic performance and good tissue compatibility, and therefore may be an alternative to other currently used valve prostheses. |
| |
Keywords: | Quality of life heart failure validation studies |
|