Abstract: | Summary Reticulocyte counting was assessed on the Coulter® STKS-2A automated blood cell counter. Using a two step procedure, blood samples were first incubated with the supravital stain new methylene blue. An acidic reagent was then added to clear the haemoglobin and any stained RNA was preserved within the cell. The cells were then analysed by measurement of volume, conductivity and light scatter (VCS). The results of 123 samples analysed on the STKS-2A were compared with those from a Toa Sysmex R-1000 reticulocyte counter. One hundred and seven samples gave no review flags and reticulocyte counts ranging from 0.5% to 22.8%, resulting in a correlation coefficient of 0.93 for the methods. Between run imprecision studies gave CVs ranging from 5.3% for a reticulocyte count of 8.7% to a CV of 16.3% for a 0.34% count. Stability studies showed insignificant changes over 72 h storage. These findings confirm that VCS technology can be adapted to provide precise and accurate routine reticulocyte analysis. |