Automatic Implantable Cardioverter/ Defibrillator (AICD) and Antitachycardia Pacemaker (Tachylog): Combined use in Ventricular Tachyarrhythmias
Authors:BERNDT LÜ  DERITZ,ULRICH GERCKENS,MATTHIAS MANZ
Affiliation:Department of Cardiology, University of Bonn, Federal Republic of Germany
Abstract:Antitachycardia pacing in ventricular tachyarrhythmias (VT) is associated with potential acceleration of VT: frequency of VT and discomfort of the patient (pt) can limit treatment with the AICD. We therefore evaluated the combined use of antitachycardia pacing and AICD in 6 of 14 patients (age 50–70. mean 60 years) with AICD implantation because of VT, which could he terminated by temporary overdrive pacing. With the interactive mode of the Tachylog, termination of VT by the pacemaker as well as by the AICD was assessed after implantation. In the automatic mode, the Tachylog functioned as a bipolar VVI device with antitachycardia burst stimulation: 2–5 stimuli, interval 260–300 ms. 1–2 interventions. During follow-up of 12±5 months, the Tachylog terminated VT reliably 20 to 327 times per patient. In three patients, burst stimulation accelerated VT, which was terminated then by the AICD discharge. Conclusion: Drug resistant ventricular tachyarrhythmias can be terminated by the Tachylog pacemaker avoiding patients' discomfort. In case of acceleration. ventricular tachyarrhythmias can be controlled by the automatic implantable cardioverter/defibrillator. A universal pacemaker should combine antitachycardia pacing with back-up defibrillation mode.
Keywords:antitachycardia pacemaker    automatic implantable cardioverter/defibrillator    ventricular tachycardia