Dosimetric comparison at FiR 1 using microdosimetry, ionisation chambers and computer simulation.
Authors:J Uusi-Simola  T Serén  T Sepp?l?  A Kosunen  I Auterinen  S Savolainen
Affiliation:Department of Physical Sciences, University of Helsinki, PO BOX 64, FI-00014 Helsinki, Finland. jouni.uusi-simola@hus.fi
Abstract:Tissue equivalent proportional counter microdosimetry has been applied in the dosimetry of epithermal neutron beams as they can provide an independent and accurate method to determine gamma ray and neutron absorbed doses. Dosimetric comparison has been performed using a tissue equivalent proportional counter, dual ionisation chambers and DORT computer code at FiR 1 boron neutron capture therapy facility in Espoo, Finland. The three methods were applied to determine neutron and gamma ray absorbed doses at 25, 40, 60 and 120 mm depths along the beam centerline in a water-filled PMMA phantom. The determined absorbed doses were found to agree within the limits of the estimated uncertainties.