Effect of contrast material on image noise and radiation dose in adult chest computed tomography using automatic exposure control: a comparative study between 16-, 64- and 128-slice CT
Authors:Paul Jijo  Schell Boris  Kerl J Matthias  Maentele Werner  Vogl Thomas J  Bauer Ralf W
Institution:a Clinic of the Goethe University, Department of Diagnostic and Interventional Radiology, Haus 23C UG, Theodor-Stern-Kai 7, 60590 Frankfurt am Main, Germany
b Goethe University, Department of Biophysics, Max von Laue-Strasse 1, 60438 Frankfurt am Main, Germany
Abstract:

Purpose

To determine the difference in radiation dose between non-enhanced (NECT) and contrast-enhanced (CECT) chest CT examinations contributed by contrast material with different scanner generations with automatic exposure control (AEC).

Methods & materials

Each 42 adult patients received a NECT and CECT of the chest in one session on a 16-, 64- or 128-slice CT scanner with the same scan protocol settings. However, AEC technology (Care Dose 4D, Siemens) underwent upgrades in each of the three scanner generations. DLP, CTDIvol and image noise were compared.

Results

Although absolute differences in image noise were very small and ranged between 10 and 13 HU for NECT and CECT in median, the differences in image noise and dose (DLP: 16-slice:+2.8%; 64-slice:+3.9%; 128-slice:+5.6%) between NECT and CECT were statistically significant in all groups. Image noise and dose parameters were significantly lower in the most recent 128-slice CT generation for both NECT and CECT (DLP: 16-slice:+35.5-39.2%; 64-slice:+6.8-8.5%).

Conclusion

The presence of contrast material lead to an increase in dose for chest examinations in three CT generations with AEC. Although image noise values were significantly higher for CECT, the absolute differences were in a range of 3 HU. This can be regarded as negligible, thus indicating that AEC is able to fulfill its purpose of maintaining image quality. However, technological developments lead to a significant reduction of dose and image noise with the latest CT generation.
Keywords:Chest CT  Automatic exposure control  16-slice CT  64-slice  CT 128-slice  CT contrast material