首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of methods for improving precision of blood pressure measurements in phase I clinical trials
Authors:Terra Steven G  Blum Robert A  Wei Greg C  Lew Robert A  Digenio Andres G  Rajman Iris  Kazierad David J
Affiliation:Pfizer Global Research & Development, 445 Eastern Point Road, BLDG 260/2505, Groton, CT 06340, USA.
Abstract:Small sample sizes are typically incorporated in early Phase I clinical studies, which may lead to insignificant changes in safety parameters such as blood pressure. Therefore, it is paramount to identify an optimal, noninvasive method of accurately measuring blood pressure and an appropriate analysis strategy yielding the smallest variability. The goals of this study were (1) to compare the variability between automated and manual blood pressure measurements, (2) to determine whether triplicate blood pressure measurements were independent of one another, and (3) to assess how the number of blood pressure readings affects variability and study sample size. Twenty healthy volunteers were enrolled in this randomized, two-way crossover study. Each subject received three incremental infusions of phenylephrine or normal saline on separate days to simulate blood pressure variability. The mean systolic blood pressure readings with the automated device were consistently higher than the manual device by 3 to 5 mmHg. Conversely, the mean diastolic blood pressure readings with the automated device were consistently 3 to 5 mmHg lower than the manual device. However, the variability and absolute change in blood pressure were essentially identical with manual and automated methods. No systematic order effects such as the first blood pressure reading always being higher were detected, suggesting that the triplicate readings were independent of one another and that an interval of 2 minutes between readings is adequate. Compared to a single measurement, collecting blood pressure in triplicate results in a 40% lower sample size needed to detect a 5-mmHg difference in systolic blood pressure.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号