首页 | 本学科首页   官方微博 | 高级检索  
     


An evaluation of a noninvasive cardiac output measurement using partial carbon dioxide rebreathing in children
Authors:Levy Richard J  Chiavacci Rosetta M  Nicolson Susan C  Rome Jonathan J  Lin Richard J  Helfaer Mark A  Nadkarni Vinay M
Affiliation:Department of Anesthesiology and Critical Care Medicine, The Children's Hospital of Philadelphia, 34th St. and Civic Center Blvd., Philadelphia, PA 19104, USA. levyri@email.chop.edu
Abstract:Cardiac output (CO) is an important hemodynamic measure that helps to guide the therapy of critically ill patients. Invasive CO assessment in infants and children is often avoided because of the inherent risks. A noninvasive CO monitor that uses partial rebreathing has been recently developed to determine CO via the Fick principle for carbon dioxide. There have been no clinical studies confirming its accuracy in pediatric patients. This is a prospective observational study of 37 children <12 yr of age who underwent cardiac catheterization. Under general anesthesia via an endotracheal tube without a leak, we made multiple CO measurements using thermodilution and compared them with noninvasively determined CO measurements. Paired measurements were analyzed for bias, precision, and correlation via Bland-Altman plot and linear regression. Noninvasive measurements showed a linear correlation with thermodilution CO assessment with an r value of 0.83 (P < 0.03). Bland-Altman analysis yielded a bias of -0.27 L/min and a precision +/-1.49 L/min. Cardiac index measurements demonstrated a decreased r value of 0.67 (P = 0.15) and a bias of -0.18 L . min(-1) . m(-2) and precision of +/-2.13 L . min(-1) . m(-2). Differences between partial rebreathing measurements and thermodilution measurements were largest in children with a body surface area of 0.6 m(2) body surface area and >300 mL tidal volume.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号