Diaton tonometry: an assessment of validity and preference against Goldmann tonometry
Authors:Doherty Mark D  Carrim Zia I  O'Neill Damian P
Institution:Department of Ophthalmology, St James's University Hospital, Leeds, UK. highburymark@yahoo.co.uk
Abstract:Background: To assess agreement between the Diaton, a new transpalpebral tonometer, and Goldmann applanation tonometry, the accepted gold standard. Design: Comparative study of two devices in a hospital setting. Participants: Two hundred and fifty‐one patients attending the eye casualty and general ophthalmology clinics at St James' University Hospital, Leeds between February and December 2009. Methods: Intraocular pressure was measured using Goldmann applanation tonometry and Diaton tonometry by one examining ophthalmologist. Patient preference for either technique was also recorded. Main Outcome Measures: Intraocular pressure measured by Diaton was compared with intraocular pressure measured by Goldmann applanation tonometry. Limits of agreement were determined using the Bland‐Altman method. Results: Two hundred and fifty right eyes underwent both Goldmann applanation tonometry and Diaton tonometry. Mean intraocular pressure was 13.8 ± 3.6 mmHg using Goldmann applanation tonometry and 13.2 ± 4.3 mmHg using Diaton tonometry. Upper and lower limits of agreement were +8.4 mmHg and ?9.6 mmHg, respectively. Order of intraocular pressure measurement and positioning did not influence limits of agreement in a clinically significant manner. Overall, more patients expressed preference for Diaton tonometry (40.2%) than Goldmann applanation tonometry (30.3%). Those aged 50 or less were more likely to prefer Diaton tonometry. Conclusions: The Diaton tonometer is portable, lightweight, user‐friendly and well tolerated by patients. However, it shows poor agreement with Goldmann applanation tonometry, thereby precluding it from being regarded as a substitute in routine clinical practice.
Keywords:Diaton  Goldmann applanation tonometry  intraocular pressure
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号