Reliability of the MicroScan WalkAway PC21 panel in identifying and detecting oxacillin resistance in clinical coagulase-negative staphylococci strains |
| |
Authors: | A. N. Olendzki E. M. Barros M. S. Laport K. R. N. dos Santos M. Giambiagi-deMarval |
| |
Affiliation: | 1. Departamento de Microbiologia Médica, Instituto de Microbiologia Paulo de Góes, Universidade Federal do Rio de Janeiro, Av. Carlos Chagas Filho, 373, Cidade Universitária, 21941-590, Rio de Janeiro, Brazil 2. Laboratório de Bacteriologia, Servi?o de Patologia Clínica, Hospital Naval Marcílio Dias, R. César Zama, 185, Lins de Vasconcelos, 20725-090, Rio de Janeiro, Brazil
|
| |
Abstract: | The purpose of this study was to determine the reliability of the MicroScan WalkAway PosCombo21 (PC21) system for the identification of coagulase-negative staphylococci (CNS) strains and the detection of oxacillin resistance. Using molecular and phenotypic methods, 196 clinical strains were evaluated. The automated system demonstrated 100 % reliability for the identification of the clinical strains Staphylococcus haemolyticus, Staphylococcus hominis and Staphylococcus cohnii; 98.03 % reliability for the identification of Staphylococcus epidermidis; 70 % reliability for the identification of Staphylococcus lugdunensis; 40 % reliability for the identification of Staphylococcus warneri; and 28.57 % reliability for the identification of Staphylococcus capitis, but no reliability for the identification of Staphylococcus auricularis, Staphylococcus simulans and Staphylococcus xylosus. We concluded that the automated system provides accurate results for the more common CNS species but often fails to accurately identify less prevalent species. For the detection of oxacillin resistance, the automated system showed 100 % specificity and 90.22 % sensitivity. Thus, the PC21 panel detects oxacillin-resistant strains, but is limited by the heteroresistance that is observed when using most phenotypic methods. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|