Affiliation:1. University of California, Berkeley, Berkeley, California, 2521 Hearst Avenue, Etcheverry Hall Rm#2121, Berkeley, CA 94709, United States;2. Stanford University, Stanford, California, Stanford University Medical Center, Stanford Medicine Outpatient Center, 450 Broadway Street, Mailcode 6342, Redwood City, CA 94063, United States;3. Dartmouth College, Hanover, New Hampshire, Thayer School of Engineering, 14 Engineering Drive, Hanover, NH 03755, United States;4. University of California, San Francisco, San Francisco, California, Department of Orthopaedic Surgery, 500 Parnassus Ave (MU320-W), San Francisco, CA 94143, United States
Abstract:Highly cross-linked ultrahigh molecular weight polyethylene (UHMWPE) is increasingly used as a bearing material in total hip replacements. Cross-linking of UHMWPE has been shown to increase wear resistance but decrease its fracture resistance. We analyzed the clinical fracture failure of four cross-linked UHMWPE total hip replacement components of four different designs via microscopic observation of the fracture surfaces, and found that in all cases fractures initiated at stress concentrations in an unsupported region of the component (termed the elevated rim). Finite element analyses (FEA) of each individual implant design were then conducted. Results from this analysis demonstrated that the predicted magnitude and orientation of maximum principal stress due to mechanical loading of the elevated rim was sufficient to propagate initiated fatigue cracks in each case. FEA also predicted that cracks may arrest after some amount of growth due to a steep stress gradient near the initiation site. Further, while anatomical positioning of the implant and material properties affect the risk of fracture, we examined whether these failures are strongly related to the notched elevated rim design feature that is common to the four failed cases presented here. We believe that cross-linked UHMWPE remains an excellent bearing material for total hip replacements but that designs employing this material should mitigate stress concentrations or other design features that increase the risk of fracture.
Keywords:Acetabular liner   UHMWPE   Notch fatigue   Fractography   Crosslinking   Fracture
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号