Abstract:The introduction of the variable-pitch feature on pulse oximeters in 1983 by the Nellcor Corporation (Hayward, CA) allowed users to rapidly detect changes in oxygen saturation by listening for changes in the pitch of the tones emitted by the pulse oximeter. A few individuals have reported that they have been unable to detect a change in pitch when oxygen saturation changes. To these individuals, the variable-pitch feature of these pulse oximeters has not been beneficial. Using the pitches from one manufacturer of oximeters, we created a computer program to simulate the pitches that accompanied various oxygen saturations. The pitches were recorded onto a tape player and played for 75 volunteer subjects unfamiliar with the pitches of a variable-pitch pulse oximeter. Of our sample, 67% were able to detect a single change in pitch corresponding to a 1% fall in oxygen saturation, and 11% of the population could not detect a change in pitch until there was a change in pitch with every beat. We suggested four alternative designs that may prove beneficial to this group of individuals.
Keywords:Measurement techniques: pulse oximetry  Equipment: pulse oximeters  Hearing