低场MR二维快速扰相梯度回波序列在颅脑检查中的应用
引用本文:胡晓云,胡春洪,王苇,张海平,宋璠.低场MR二维快速扰相梯度回波序列在颅脑检查中的应用[J].中国医学影像技术,2005,21(10):1579-1581.
作者姓名:胡晓云  胡春洪  王苇  张海平  宋璠
作者单位:1. 苏州大学附属第一医院影像科,江苏,苏州,215006
2. 东南大学附属扬州市第一人民医院影像科,江苏,扬州,225001
3. 江苏省高邮市人民医院放射科,江苏,高邮,225600
摘    要:目的观察快速扰相梯度回波(2D FSPGR)在颅脑低场MR检查中的应用价值.方法随机对70例颅脑用翻转角(FA)为70°的2D FSPGR(2D FSPGR 70°)获得T1加权图像,与常规自旋回波(SE)T1WI作对照,比较轴向图像的成像速度、血管显示、清晰度、增强效果、灰白质对比等方面,计算相关的质量控制参数,包括信噪比(SNR)、对比度(C)及对比噪声比(CNR).结果2D FSPGR 70°的图像具有成像快速、血管显示为高信号、增强效果佳等特点.2D FSPGR与SET1WI的SNR、C及CNR分别为81.98、0.116、6.58和87.15、0.117、7.18,二者基本相当.结论2D FSPGR 70°在颅脑低场MR检查中具有许多SE T1WI无法比拟的优势,可以代替SE T1WI或作为T1加权的优先替补序列.

关 键 词:磁共振成像  快速扰相梯度回波  
文章编号:1003-3289(2005)10-1579-03
收稿时间:2005-05-05
修稿时间:2005-07-29

Low field MR 2D fast spoiled gradient echo sequence in application of cerebral scan
HU Xiao-yun,HU Chun-hong,WANG Wei,ZHANG Hai-ping and SONG Fan.Low field MR 2D fast spoiled gradient echo sequence in application of cerebral scan[J].Chinese Journal of Medical Imaging Technology,2005,21(10):1579-1581.
Authors:HU Xiao-yun  HU Chun-hong  WANG Wei  ZHANG Hai-ping and SONG Fan
Institution:Department of Radiology, the First Affiliated Hospital of Suzhou University, Suzhou 215006, China;Department of Radiology, the First Affiliated Hospital of Suzhou University, Suzhou 215006, China;Department of Radiology, the Affiliated Yangzhou Hospital of Southeast University, Yangzhou 225001, China;Department of Radiology, Gaoyou People's Hospital of Jiangsu, Gaoyou 225600, China;Department of Radiology, Gaoyou People's Hospital of Jiangsu, Gaoyou 225600, China
Abstract:Objective To investigate the value of two-dimensional fast spoiled gradient echo (2D FSPGR) sequence in the application of cerebral scan of low field MRI. Methods By 0.2 T MR scanner, 70 cases underwent both 2D FSPGR and SE T1WI randomly, and the scan-time, vessel display, definition, enhanced effect and grey-white contrast of 2D FSPGR and SE T1WI were compared. Results Two-dimensional FSPGR was characterized by short scan-time, high-signal vessel and excellent enhanced effect, with the results of SNR (81.98), C (0.116), CNR (6.58) approximated to those of SE T1WI (SNR 87.15, C 0.117, CNR 7.18). Conclusion Two-dimensional FSPGR is superior to SE T1WI in many aspects with high value in the clinical application. It may replace SE T1WI or can be taken as a supplementary sequence for T1WI.
Keywords:Magnetic resonance imaging  Fast spoiled gradient echo~ Brain