A megavoltage CT scanner for radiotherapy verification.
Authors:D G Lewis  W Swindell  E J Morton  P M Evans  Z R Xiao
Affiliation:Joint Department of Physics, Royal Marsden Hospital and Institute of Cancer Research, Sutton, Surrey, UK.
Abstract:We have further developed a system for generating megavoltage CT images immediately prior to the administration of external beam radiotherapy. The detector is based on the scanner of Simpson (Simpson et al 1982)--the major differences being a significant reduction in dose required for image formation, faster image formation and greater convenience of use in the clinical setting. Attention has been paid to the problem of ring artefacts in the images. Specifically, a Fourier-space filter has been applied to the sinogram data. After suitable detector calibration, it has been shown that the device operates close to its theoretical specification of 3 mm spatial resolution and a few percent contrast resolution. Ring artefacts continue to be a major source of image degradation. A number of clinical images have been presented. The next stage of this work is to use the system to make clinical measurements of patient set-up inaccuracies building on our work making such measurements from digital portal images (Evans et al 1992).