Validation of noninvasive pulse contour cardiac output using finger arterial pressure in cardiac surgery patients requiring fluid therapy
Authors:Charlotte Hofhuizen  Benno Lansdorp  Johannes G. van der Hoeven  Gert-Jan Scheffer  Joris Lemson
Affiliation:1. Radboud University Nijmegen Medical Centre, Department of Anesthesiology, PO Box 9101 6500 HB Nijmegen, The Netherlands;2. Radboud University Nijmegen Medical Centre, Department of Intensive Care Medicine, 6500 HB Nijmegen, The Netherlands;3. University of Twente, MIRA Institute for Biomedical Technology and Technical Medicine, 7500 AE Enschede, The Netherlands
Abstract:

Introduction

Nexfin (Edwards Lifesciences, Irvine, CA) allows for noninvasive continuous monitoring of blood pressure (ABPNI) and cardiac output (CONI) by measuring finger arterial pressure (FAP). To evaluate the accuracy of FAP in measuring ABPNI and CONI as well as the adequacy of detecting changes in ABP and CO, we compared FAP to intra-arterially measured blood pressure (ABPIA) and transpulmonary thermodilution (COTD) in postcardiac surgery patients during a fluid challenge (FC).

Methods

Twenty sedated patients post cardiac surgery were included, and 28 FCs were performed. Measurements of ABP and CO were simultaneously collected before and after an FC, and we compared CO and blood pressure.

Results

Finger arterial pressure was obtainable in all patients. When comparing ABPNI with ABPIA, bias was 2.7 mm Hg (limits of agreement [LOA], ± 22.2), 4.9 mm Hg (LOA, ± 13.6), and 4.2 mm Hg (LOA, ± 13.7) for systolic, diastolic, and mean arterial pressure, respectively. Concordance between changes in ABPNI and ABPIA was 100%. Mean bias between CONI and COTD was − 0.26 (LOA, ± 2.2), with a percentage error of 38.9%. Concordance between changes in CONI vs COTD and was 100%.

Conclusion

Finger arterial pressure reliably measures ABP and adequately tracks changes in ABP. Although CONI is not interchangeable with COTD, it follows changes in CO closely.
Keywords:Arterial pressure   Measurement   Equipment   Nexfin   Fluid therapy   Cardiac output   Measurement