Current Data Do Not Support Routine Use of Patient-Specific Instrumentation in Total Knee Arthroplasty
Authors:Pramod B. VoletiMathew J. Hamula  BS  Keith D. BaldwinGwo-Chin Lee  MD
Affiliation:Department of Orthopaedic Surgery, Hospital of the University of Pennsylvania, Philadelphia, Pennsylvania
Abstract:The purpose of this systematic review and meta-analysis is to compare patient-specific instrumentation (PSI) versus standard instrumentation for total knee arthroplasty (TKA) with regard to coronal and sagittal alignment, operative time, intraoperative blood loss, and cost. A systematic query in search of relevant studies was performed, and the data published in these studies were extracted and aggregated. In regard to coronal alignment, PSI demonstrated improved accuracy in femorotibial angle (FTA) (P = 0.0003), while standard instrumentation demonstrated improved accuracy in hip-knee-ankle angle (HKA) (P = 0.02). Importantly, there were no differences between treatment groups in the percentages of FTA or HKA outliers (>3 degrees from target alignment) (P = 0.7). Sagittal alignment, operative time, intraoperative blood loss, and cost were also similar between groups (P > 0.1 for all comparisons).
Keywords:patient-specific instrumentation   patient-specific guides   custom cutting guides   total knee arthroplasty instrumentation   primary total knee arthroplasty
本文献已被 ScienceDirect 等数据库收录!