首页 | 本学科首页   官方微博 | 高级检索  
     


Development of a New Hollow Fiber Silicone Membrane Oxygenator: In Vitro Study
Authors:Shinji   Kawahito,Tomohiro   Maeda,Tadashi   Motomura,Tamaki   Takano,Kenji   Nonaka,Joerg   Linneweber,Minoru   Mikami,Seiji   Ichikawa,Masaki   Kawamura,Julie   Glueck,&dagger  Koshiro   Sato, Yukihiko   Nosé  
Affiliation:Baylor College of Medicine, Michael E. DeBakey Department of Surgery, Houston, Texas, U.S.A. kawahito@clin.med.tokushima-u.ac.jp
Abstract:An experimental silicone hollow fiber membrane oxygenator for long-term extracorporeal membrane oxygenation (ECMO) was developed in our laboratory using an ultrathin silicone hollow fiber. However, the marginal gas transfer performances and a high-pressure drop in some cases were demonstrated in the initial models. In order to improve performance the following features were incorporated in the most recent oxygenator model: increasing the fiber length and total surface area, decreasing the packing density, and modifying the flow distributor. The aim of this study was to evaluate the gas transfer performances and biocompatibility of this newly improved model with in vitro experiments. According to the established method in our laboratory, in vitro studies were performed using fresh bovine blood. Gas transfer performance tests were performed at a blood flow rate of 0.5 to 6 L/min and a V/Q ratio (V = gas flow rate, Q = blood flow rate) of 2 and 3. Hemolysis tests were performed at a blood flow rate of 1 and 5 L/min. Blood pressure drop was also measured. At a blood flow rate of 1 L/min and V/Q = 3, the O2 and CO2 gas transfer rates were 72.45 +/- 1.24 and 39.87 +/- 2.92 ml/min, respectively. At a blood flow rate of 2 L/min and V/Q = 3, the O2 and CO2 gas transfer rates were 128.83 +/- 1.09 and 47.49 +/- 5.11 ml/min. Clearly, these data were superior to those obtained with previous models. As for the pressure drop and hemolytic performance, remarkable improvements were also demonstrated. These data indicate that this newly improved oxygenator is superior to the previous model and may be clinically acceptable for long-term ECMO application.
Keywords:Hollow fiber    Silicone membrane    Oxygenator    Extracorporeal membrane oxygenation
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号