首页 | 本学科首页   官方微博 | 高级检索  
     


Development and validation of a semi-automated method for L-dopa and dopamine in rat plasma using electrospray LC/MS/MS
Authors:Li W  Rossi D T  Fountain S T
Affiliation:Department of Pharmacokinetics, Dynamics, Metabolism, Pfiezer Global Research and Development, Ann Arbor, MI 48105, USA.
Abstract:A semi-automated alumina-based extraction method for the determination of L-dopa and dopamine in plasma using liquid chromatography mass spectrometry was validated. The method exploited the use of a Tomtec Quadra 96 liquid handing robot to expedite aluminum oxide extraction for sample clean up. Two 96-well sample plates can be processed in less than 2 h and extracts, collected in a 96-well plate format, can be directly injected onto the ESI/LC/MS/MS instrumentation. Chromatographic separation of the analytes was performed on a reverse-phase ODS column (TosoHaas ODS-80) with a mobile phase of acetonitrile/0.1% formic acid (5/95 v/v) at a flow rate of 0.22 ml/min. Analytes were detected by a triple-quadruple mass spectrometer equipped with an electrospray ionization source (ESI). Recoveries were evaluated for a number of pH modifiers and elution solvents. Under optimized conditions, the mean recoveries of L-dopa and dopamine were 56 and 67%, respectively. Intra-run and inter-run precision, calculated as percent relative standard deviation of replicate quality controls, was in the range of 1.45-10.8% for both L-dopa and dopamine. Intra-run and inter-run accuracy, calculated as percent error, was in the range -2.5 to 6.69% for both analytes. The limit of quantitaiton was 2.5 ng/ml for both L-dopa and dopamine when 100 microl of plasma was extracted. The method is simple, rapid, accurate and suitable for the quantification of L-dopa and dopamine in plasma or other biological fluid samples from clinical, preclinical, or pharmacological studies.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号