Calibrating automatic exposure control devices for digital radiography |
| |
Authors: | Doyle P Martin C J |
| |
Affiliation: | Health Physics, Department of Clinical Physics and Bio-Engineering, Gartnavel Royal Hospital, Glasgow G12 OXH, UK. philip.doyle@northglasgow.scot.nhs.uk |
| |
Abstract: | The energy responses of digital radiography detectors differ from those of screen-film systems. To provide a consistent level of image quality at different tube potentials automatic exposure control (AEC) devices must be calibrated to suit the energy response of the image receptor with which they are intended for use. AEC calibration for digital radiography systems requires an alternative parameter to optical density, ideally one related to the quality of a digital image. Energy responses of computed radiography (CR) and indirect digital radiography (IDR) image receptors have been calculated, and compared with those for screen-film systems. Practical assessments of the relative sensitivities of a CR detector made using the detector dose indicator (DDI), pixel value and signal-to-noise ratio showed similar variations with tube potential. The DDI has been used to determine the correct kV compensation curve required to calibrate the AECs for the loss in detector sensitivity with tube potential. AECs are set up relative to a predetermined air kerma incident on the detector at 80 kV for CR and IDR systems using this curve and the method used is described. Factors influencing the calibration of AECs for digital radiography including techniques, types of phantom and contributions from scatter are reviewed, and practical methods recommended for use. |
| |