首页 | 本学科首页   官方微博 | 高级检索  
     


Curing efficiency and ocular hazards of dental photopolymerization sources
Authors:Wayne D. Cook
Affiliation:

Australian Dental Standards Laboratory, 240 Langridge Street, Abbotsford, Victoria 3067, Australia

Abstract:The spectral distributions of a range of dental photocuring sources were measured at the exit window and at a distance of 10 cm. The former enabled the evaluation of a newly proposed photocuring efficiency index which correlates well with the depth of cure of the photopolymerized resins, thus providing a basis for the comparison of different photocuring sources. The spectral irradiante of the sources obeyed the inversesquare law, allowing a comparison with the ACGIH threshold limit values. According to these criteria, no ocular hazard is posed to the patient or clinician by u.v.-A or u.v.-B radiation nor to the patient by the visible light when momentarily exposed to the sources. Similarly the ACGIH criterion indicates that the clinician does not risk chorioretinal injury provided the exposure is restricted to less than 140 s in a 3 h period.
Keywords:Dental materials   photopolymerization   depth of cure   composite resins   ocular hazards   photocuring sources
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号