首页 | 本学科首页   官方微博 | 高级检索  
检索        


Laboratory performance of alternating pressure air mattresses component and sequelae
Authors:Bain Duncan
Institution:Duncan Bain Consulting , Herfordshire.
Abstract:The performance of three different alternating pressure air mattresses with different geometries of air cell were compared (Nimbus 3, Heritage, Tamora Plus), using simple performance indices based on pressure mapping. The aim of this study was to examine the effect on performance of elevating the backrest and thigh section of the bed into sitting position. Ten healthy volunteers of various sizes were pressure-mapped over the full pressure cycle on three alternating pressure air mattresseses with differing cell geometries. This was then repeated with the beds profiled to a sitting position. Performance of the alternating pressure air mattresses in terms of their ability to redistribute pressure dynamically was assessed in the different positions. The different alternating pressure air mattresses performed similarly with the bed in the lying flat position, but smaller cells appeared to be more effective in the sitting position. A conclusion was made that cell geometry may have an effect on the ability of the mattress to achieve alternating behaviour in the sitting position.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏