首页 | 本学科首页   官方微博 | 高级检索  
检索        


Dose-rate scaling factor estimation of THOR BNCT test beam.
Authors:F Y Hsu  C J Tung  J C Chen  Y L Wang  H C Huang  R G Zamenhof
Institution:Yuanpei University of Science and Technology, 306 Yuanpei Street, Hsinchu 300, Taiwan. fyhsu@mail.yust.edu.tw
Abstract:In 1998, an epithermal neutron test beam was designed and constructed at the Tsing Hua Open-Pool Reactor (THOR) for the purpose of preliminary dosimetric experiments in boron neutron capture therapy (BNCT). A new epithermal neutron beam was designed at this facility, and is currently under construction, with clinical trials targeted in late 2004. Depth dose-rate distributions for the THOR BNCT test beam have been measured by means of activation foil and dual ion chamber techniques. Neutron and structure-induced gamma spectra measured at the test beam exit were configured into a source function for the Monte Carlo-based treatment planning code NCTPlan. Dose-rate scaling factors (DRSFs) were determined to normalize computationally derived dose-rate distributions with experimental measurements in corresponding mathematical and physical phantoms, and to thus enable accurate treatment planning using the NCTPlan code. A similar approach will be implemented in characterizing the new THOR epithermal beam in preparation for clinical studies. This paper reports the in-phantom calculated and experimental dosimetry comparisons and derived DRSFs obtained with the THOR test beam.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏