Comparison of daily megavoltage electronic portal imaging or kilovoltage imaging with marker seeds to ultrasound imaging or skin marks for prostate localization and treatment positioning in patients with prostate cancer
Authors:Serago Christopher F  Buskirk Steven J  Igel Todd C  Gale Ashley A  Serago Nicole E  Earle John D
Institution:Department of Radiation Oncology, Mayo Clinic, Jacksonville, FL 32224, USA. serago.christopher@mayo.edu
Abstract:PURPOSE: To compare the accuracy of imaging modalities, immobilization, localization, and positioning techniques in patients with prostate cancer. METHODS AND MATERIALS: Thirty-five patients with prostate cancer had gold marker seeds implanted transrectally and were treated with fractionated radiotherapy. Twenty of the 35 patients had limited immobilization; the remaining had a vacuum-based immobilization. Patient positioning consisted of alignment with lasers to skin marks, ultrasound or kilovoltage X-ray imaging, optical guidance using infrared reflectors, and megavoltage electronic portal imaging (EPI). The variance of each positioning technique was compared to the patient position determined from the pretreatment EPI. RESULTS: With limited immobilization, the average difference between the skin marks' laser position and EPI pretreatment position is 9.1 +/- 5.3 mm, the average difference between the skin marks' infrared position and EPI pretreatment position is 11.8 +/- 7.2 mm, the average difference between the ultrasound position and EPI pretreatment position is 7.0 +/- 4.6 mm, the average difference between kV imaging and EPI pretreatment position is 3.5 +/- 3.1 mm, and the average intrafraction movement during treatment is 3.4 +/- 2.7 mm. For the patients with the vacuum-style immobilization, the average difference between the skin marks' laser position and EPI pretreatment position is 10.7 +/- 4.6 mm, the average difference between kV imaging and EPI pretreatment position is 1.9 +/- 1.5 mm, and the average intrafraction movement during treatment is 2.1 +/- 1.5 mm. CONCLUSIONS: Compared with use of skin marks, ultrasound imaging for positioning provides an increased degree of agreement to EPI-based positioning, though not as favorable as kV imaging fiducial seeds. Intrafraction movement during treatment decreases with improved immobilization.
Keywords:Prostate cancer  Localization  Imaging  Fiducial markers  Radiotherapy
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号