首页 | 本学科首页   官方微博 | 高级检索  
     


Effects of tongue position and base circle diameter on the performance of a centrifugal blood pump
Authors:Wong Yew-Wah  Chan Weng-Kong  Hu Wei
Affiliation:School of Mechanical and Aerospace Engineering, Nanyang Technological University, Republic of Singapore.
Abstract:This article presents numerical investigations of the effect of radial gap and volute tongue position on the circumferential pressure distribution and the magnitude of resulting imbalanced radial force. A series of volute models was designed using the constant mean velocity method. The results indicate that a radial clearance of 10% is a good practical value that gives a relatively high head across the pump for a small radial force. The results show that the tongue position at 30 degrees gives the lowest radial force and pressure head. The tongue position at 15 degrees appears to give the best compromise results producing a generated head only 5% less than the maximum value while the radial force is about 22% less than the maximum force.
Keywords:Volute radial clearance    Volute tongue position    Centrifugal blood pump
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏