首页 | 本学科首页   官方微博 | 高级检索  
     

血液透析设备检测装置溯源校准方法分析
引用本文:杨冬,孙志辉,李咏雪,葛剑徽. 血液透析设备检测装置溯源校准方法分析[J]. 医疗卫生装备, 2014, 0(6): 22-24
作者姓名:杨冬  孙志辉  李咏雪  葛剑徽
作者单位:[1]全军医学计量测试研究中心,北京100071 [2]南京军区联勤部药品仪器检验所,南京210002
基金项目:军事医学计量专项科研课题(2011-JL2-059)
摘    要:目的:研究建立血液透析设备检测装置的溯源校准方案,完善血液透析设备检测装置的溯源体系建设。方法:从血液透析机检测装置的溯源校准需求及标准依据入手,开展检测试验,详尽分析血液透析机检测装置的电导率、温度、pH值、压力、流量等参数的校准方法结果:通过试验,修正了各参数的校准方法,制订了可满足各参数校准需求的溯源校准方案结论:该溯源校准方案为血液透析检测仪校准方法的评定及开展相关试验研究提供了理论支持,为军队建立血液透析设备检测技术规范与质量控制规章制度、深入推广血液透析设备质量控制检测工作打下了坚实的基础、

关 键 词:血液透析  检测装置  溯源校准  检测技术

Method for traceability and calibration of testing device of dialysis equipment
YANG Dong,SUN Zhi-hui,LI Yong-xue,GE Jian-hui. Method for traceability and calibration of testing device of dialysis equipment[J]. Chinese Medical Equipment Journal, 2014, 0(6): 22-24
Authors:YANG Dong  SUN Zhi-hui  LI Yong-xue  GE Jian-hui
Affiliation:1. Research Center for Military Medical Metrology of the PLA, Beijing 100071, China; 2. Institute of Drug and Instrument Control, Joint Logisties Department of Nanjing Military Area Command, Nanjing 210002, China)
Abstract:Objective To establish a seheme for the traceability and ealibration of the testing deviee of the dialysis equipment in order to complete its traceability system. Methods The standard requirements and bases of the traceability were analyzed, and then the calibration method for such parameters of the testing device were explored as the conductivity, temperature, pH, pressure and flow. Results The calibration method of the parameters was corrected, and a traceability and ealila'ation scheme was prepared to meet the standard requirements of the parameters. Conclusion The scheme lays a foundation for the evaluation of the ealibration of the testing device of the dialysis equipment as well as the preparation of related mililary technical specifications and quality control regulations.
Keywords:dialysis  testing device  traceability and calibration  detection technology