首页 | 本学科首页   官方微博 | 高级检索  
检索        


Electronic working length determination in primary teeth by ProPex and Digital Signal Processing
Authors:Nelson-Filho Paulo  Lucisano Marcela Pacífico  Leonardo Mário Roberto  da Silva Raquel Assed Bezerra  da Silva Léa Assed Bezerra
Institution:1. Department of Paediatric Clinic, Preventive and Community Dentistry, School of Dentistry of Ribeirão Preto, University of São Paulo, Ribeirão Preto, SP, Brazil;2. Private Practice, Ribeirão Preto, SP, Brazil
Abstract:The purpose of this study was to evaluate the accuracy of electronic apex locators Digital Signal Processing (DSP) and ProPex, for root canal length determination in primary teeth. Fifteen primary molars (a total of 34 root canals) were divided into two groups: Group I - without physiological resorption (n = 16); and Group II - with physiological resorption (n = 18). The length of each canal was measured by introducing a file until its tip was visible and then it was retracted 1 mm. For electronic measurement, the devices were set to 1 mm short of the apical resorption. The data were analysed statistically using the intraclass correlation coefficient (ICC). Results showed that the ICC was high for both electronic apex locators in all situations - with (ICC: DSP = 0.82 and Propex = 0.89) or without resorption (ICC: DSP = 0.92 and Propex = 0.90). Both apex locators were extremely accurate in determining the working length in primary teeth, both with or without physiological resorption.
Keywords:apex locator  Digital Signal Processing  electronic working length  odontometry  ProPex
本文献已被 PubMed 等数据库收录!