首页 | 本学科首页   官方微博 | 高级检索  
     


Convection warmers — not just hot air
Authors:M. S. Avidan,N. Jones,R. Ing,M. Khoosal,C. Lundgren,&   D. F. Morrell
Affiliation:;Department of Anaesthetics, Area 361, Johannesburg Hospital, Private Bag X39, Johannesburg 2000, South Africa,;Department of Microbiology, South African Institute for Medical Research, PO Box 1038, Johannesburg 2000, South Africa
Abstract:We sought to determine whether the forced air convection warmers (nine Bair Huggers, Augustine Medical, and one Warm Touch, Mallinkrodt Medical) used in our operating theatres could be a source of microbial pathogens. Agar plates were placed directly in the air stream of the warmers. Four of these grew potentially pathogenic organisms. When the warmers were set to blow through perforated blankets, no growth occurred. Three of the warmers were swabbed and sites of colonisation were found in their hoses. After fixing a microbial filter to the end of the hose, organisms were no longer detectable. We conclude that these warming devices are a potential source of nosocomial infection. They should only be used in conjunction with perforated blankets, should have their microbial filters changed regularly and their hoses sterilised. The inclusion of a microbial filter into the nozzle of the hose could be incorporated into the design of the warmer.
Keywords:Equipment   temperature blankets    Infection    Hypothermia