One-month biocompatibility evaluation of the pediatric TinyPump in goats
Authors:Ando Yusuke  Kitao Takashi  Nagaoka Eiki  Kimura Taro  Yokoyama Yoshimasa  Yoshikawa Masaharu  Tominaga Ryuji  Takatani Setsuo
Affiliation:Department of Artificial Organs, Institute of Biomaterials and Bioengineering, Tokyo Medical and Dental University, Tokyo, Japan.
Abstract:The TinyPump is an extracorporeal, magnetically driven centrifugal blood pump with its impeller suspended magnetically and hydrodynamically to provide short-term mechanical circulatory support for children and infants. We have previously demonstrated that the in vivo experiments of the experimental TinyPump showed acceptable stable performance at pump flows averaging around 1.0 L/min with low hemolytic and thrombogenic properties for up to 2 weeks. We present here the 1-month in vivo evaluation of the TinyPump, whose design was modified further for more durable operation. The pump was implanted as a left ventricular assist device in five goats (12.5-26.7 kg), with inflow inserted into the left ventricular apex and outflow anastomosed to the descending aorta. Five animals were supported for 110 pump days, with mean pump flow of 1.19 ± 0.03 L/min at a pump speed of 2679 ± 97 rpm. Two animals reached the scheduled end point of 30 days without device failure, and mean plasma-free hemoglobin was 1.7 ± 0.8 mg/dL. Hematologic and biochemical data of these two animals showed no evidence of cardiovascular, renal, or hepatic dysfunction. Although further experiments are needed, the modified TinyPump offers promise as a short-term mechanical circulatory support device in pediatric population.
Keywords:Pediatric ventricular assist device  Mechanical circulatory support  Centrifugal blood pump  Hydrodynamic bearing  2‐Methacryloyloxyethyl phosphorylcholine polymer coating
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号