弹性应变率比值与BI-RADS标准鉴别诊断乳腺肿块的对比研究
引用本文:年静,郑慧,张学珍,高燕燕. 弹性应变率比值与BI-RADS标准鉴别诊断乳腺肿块的对比研究[J]. 安徽医科大学学报, 2015, 0(1)
作者姓名:年静  郑慧  张学珍  高燕燕
作者单位:安徽医科大学第一附属医院超声科,合肥,230022
基金项目:安徽省高校自然科学基金重点项目
摘    要:目的:探讨弹性应变率比值( SR)诊断乳腺良恶性肿块的最佳界点,对比分析SR与常规超声乳腺影像报告和数据系统( BI-RADS)标准对乳腺肿块的鉴别诊断价值。方法对94例患者共计101个乳腺肿块进行 SR 测定和 BI-RADS标准分级,以病理结果为标准,构建SR的受试者工作特征( ROC)曲线,确定其诊断乳腺肿块良恶性的最佳界点,并比较SR与BI-RADS标准的诊断价值。结果 SR的ROC曲线下面积(AUC)为0.933,最佳诊断界点为3.03,以SR≥3.03诊断为恶性, SR <3.03诊断为良性;良性组 SR 为(1.96±0.88),恶性组SR为(4.74±2.22),两组比较差异有统计学意义( P<0.01)。 SR和BI-RADS标准诊断乳腺肿块的敏感度、特异度、准确度、阳性预测值、阴性预测值分别为92.3%、88.7%、90.1%、83.7%、94.8%及71.8%、87.1%、81.2%、77.8%、83.1%, SR 的诊断敏感度高于 BI-RADS标准( P<0.05)。结论作为一种半定量检查方法, SR比BI-RADS标准具有更高的敏感度,可以提高乳腺恶性肿块的检出率。

关 键 词:弹性应变率比值  BI-RADS标准  乳腺疾病

Comparative study of elastic strain ratio and BI-RADS standard in differential diagnosis of breast lesions
Abstract:Objective To investigate the optimal points of elastic strain ratio ( SR) in diagnosis of benign and ma-lignant breast lesions and compare the diagnostic value of SR with BI-RADS standard. Methods 94 patients with 101 breast lesions were included in the study. The lesions were scored with BI-RADS standard and SR respective-ly. The pathologic diagnosis was the gold standard. Constructed the receiver operating characteristic ( ROC) curve and determined the optimal points of SR in the diagnosis of malignant breast lesions. Compared the diagnostic value of SR with BI-RADS standard. Results The area under the curve of SR was 0. 933,the cut-off point for strain ratio was 3. 03. The strain ratio of benign group was (1. 96 ± 0. 88),the malignant group was (4. 74 ± 2. 22),there was significant difference between two groups(P<0. 01). The sensitivity,specificity,accuracy,positive predictive value , negative predictive value of strain ratio and Conventional ultrasound were 92. 3%,88. 7%,90. 1%,83. 7%, 94. 8% and 71. 8%,87. 1%,81. 2%,77. 8%,83. 1%. Sensitivity of strain ratio was better than BI-RADS standard (P<0. 05). Conclusion As a semi quantitative method,SR is more sensitive than BI-RADS standard. It can im-prove the detection rate of malignant breast lesions.
Keywords:elastic strain ratio  BI-RADS standard  breast disease
本文献已被 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号