首页 | 本学科首页   官方微博 | 高级检索  
检索        

磁性固位体引起MRI伪影的研究
引用本文:魏斌,张富强,余强.磁性固位体引起MRI伪影的研究[J].口腔颌面修复学杂志,2000,1(4):200-202,256.
作者姓名:魏斌  张富强  余强
作者单位:上海第二医科大学
摘    要:目的:观察磁体 衔铁以及衔铁引起MRI伪影的差别。方法:以三种磁性固位体产品为样本,在0.15Tesla的永磁型磁共振仪上,分别进行体外实验和口腔内实验,对MRI伪影的最大径进行测量和分析。结果:衔铁引起伪影的最大径为91.9mm,磁体 衔铁引起伪影的最大径为145.3mm。结论:衔铁引起伪影的范围明显小于衔铁 磁体,其最大径之间存在显著差异(P<0.05)。

关 键 词:磁性固位体  磁共振成像检查  伪影  衔铁

The study of the MRI artifacts caused by magnetic attachments
WEI Bin,ZHANG Fu-qiang,YU Qiang.The study of the MRI artifacts caused by magnetic attachments[J].Chinese Journal of Prosthodontics,2000,1(4):200-202,256.
Authors:WEI Bin  ZHANG Fu-qiang  YU Qiang
Abstract:Objective: The aim of this research was to examine the difference of artifacts between whole magnetic attachment (magnetic assembly and keeper) and keeper upon MRI. Method: Three samples were imaged in a water phantom and a living human subject with 0.15 Tesla magnetic resonance devices. Results were analyzed and the maximum diameters of artifacts were measured. Results: The maximum diameter of the artifact caused by the whole magnetic attachment is 145.3mm, and that caused by keeper is 91.9mm. Conclusions: The scope of artifact caused by keeper is smaller than that caused by the whole magnetic attachment. The difference is statistically significant (p<0.05).
Keywords:magnetic attachment  MRI  artifact