Relation of stent design and stent surface material to subsequent in-stent intimal hyperplasia in coronary arteries determined by intravascular ultrasound
Authors:Hoffmann Rainer  Mintz Gary S  Haager Philipp K  Bozoglu Togul  Grube Eberhard  Gross Michael  Beythien Christian  Mudra Harald  vom Dahl Jürgen  Hanrath Peter
Affiliation:

a Medical Clinic I, University RWTH Aachen, Aachen, Germany

b Cardiovascular Research Foundation, New York, New York, USA

c Department of Cardiology, Heart Center Siegburg, Siegburg, Germany

d Department of Cardiology, University Clinic Charite, Berlin, Germany

e Department of Cardiology, University Hamburg, Hamburg, Germany

f Department of Cardiology, Krankenhaus Neuperlach, Munich, Germany

Abstract:A variety of different stent designs and coatings have become available. This study sought to determine the impact of stent design and gold-coating of stents on intimal hyperplasia (IH) in human atherosclerotic coronary arteries in relation to known predictors of restenosis. Angiographic and intravascular ultrasound (IVUS) studies were performed at 6-month follow-up on 311 native coronary lesions of 311 patients treated with 99 Multi-Link stents, 74 InFlow steel stents, 73 InFlow gold-coated stents, 41 Palmaz-Schatz stents, 12 NIR steel stents, and 12 gold-coated NIR Royal stents. Lumen and stent cross-sectional area (CSA) were measured at 1-mm axial increments. Mean IH CSA (stent CSA − lumen CSA) and mean IH thickness were calculated and averaged over the total stent length. IVUS demonstrated different levels of IH for the 6 stents. Mean IH thickness ranged from 0.20 ± 0.13 mm for Multi-Link stents to 0.43 ± 0.14 mm for InFlow goal-coated stents (p <0.001). Multivariate analysis proved non–Multi-Link stent design (odds ratio 3.45, 95% confidence intervals 1.13 to 11.11, p <0.034) and gold coating (odds ratio 3.78, 95% confidence intervals 1.88 to 7.54, p <0.001) to be the only independent predictors of IH thickness >0.3 mm. In conclusion, stent design and surface material have an important impact on the IH response to stents implanted in human coronary arteries. However, the differences in IH thickness between the analyzed stents were relatively small compared with the absolute lumen dimensions.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号