间接膀胱测压法不同调零点对腹腔内压测量的影响
引用本文:郭汉画,李小艳,王健,王芳芳. 间接膀胱测压法不同调零点对腹腔内压测量的影响[J]. 现代医学, 2014, 0(4): 382-384
作者姓名:郭汉画  李小艳  王健  王芳芳
作者单位:[1]广东省中医院重症医学科,广东广州510120 [2]广州市中医医院检验科,广东广州510130
基金项目:广州市中医药科技项目(20132A011008)
摘    要:目的:探讨间接膀胱测压法的不同调零点对测量腹腔内压力的影响,寻求能较好反映腹腔内压的最佳调零点,以指导腹腔内压的准确测量。方法:47例ICU需进行腹腔内压力测量的患者,运用间接膀胱测压法分别以腋中线与趾骨联合处为调零点进行测量,比较两者测得的腹内压力。结果:以腋中线调零点测得的腹内压均数为(18.55±1.91)mmHg,以耻骨联合处调零点测得的腹内压均数为(16.08±0.98)mmHg,两者之间差值95%置信区间为(2.08~2.86)mmHg,差异有统计学意义(P〈0.05)。结论:选择不同的调零点会对腹内压的测量造成影响,推荐以腋中线作为腹腔内压测量的调零点。

关 键 词:腹腔内压  测量  膀胱压  耻骨联合  腋中线  调零点

Impacts of different zero reference positions of indirect bladder pressure measurements on intra-abdominal pressure
GUO Han-hua,LI Xiao-yan,WANG Jian,WANG Fang-fang. Impacts of different zero reference positions of indirect bladder pressure measurements on intra-abdominal pressure[J]. Modern Medical JOurnal, 2014, 0(4): 382-384
Authors:GUO Han-hua  LI Xiao-yan  WANG Jian  WANG Fang-fang
Affiliation:1. Intensive Unit Care, Guangdong Provincial Hospital of TCM, Guangzhou 510120, China; 2. Department of Laboratory Medicine, Guangzhou Hospital of TCM , Guangzhou 510130, China )
Abstract:Objective: To investigate the impacts of different zero reference positions of indirect bladder pressure measurements on measuring intra-abdominal pressure( IAP),and look for the optimal position to reflect the IAP. Methods: Forty-seven cases of patients in ICU in need of measuring IAP were involved in this study,we used indirect bladder manometry to measure IAP,and the pressure was measured at midaxillary line and symphysis pubis respectively,and we compared the IAP at these two zero-setting places. Results: IAP measured at midaxillary line was( 18. 55 ± 1. 91) mmHg,the IAP measured at the level of the symphysis pubis was( 16. 08 ± 0. 98) mmHg, the difference between 95% confidence interval was( 2. 08-2. 86) mmHg,and it was statistically significant( P〈0. 05). Conclusion: Different zero reference levels can affect the IAP measurement,midaxillary line should be recommended as optimal zero reference position.
Keywords:intra-abdominal pressure  measurement  bladder pressure  symphysis pubis  midaxillary line  zero reference position