首页 | 本学科首页   官方微博 | 高级检索  
     


Effect of piriformis versus trochanteric starting point on fixation stability of short intramedullary reconstruction nails
Authors:Su E T  DeWal H  Sanders R  Kummer F J  Mujtaba M  Koval K J
Affiliation:NYU-Hospital for Joint Diseases Department of Orthopaedic Surgery, New York, New York 10003, USA.
Abstract:Recently, a new, shorter IM nail using two 6 mm reconstruction screws for proximal fixation was introduced in two versions for femoral insertion: piriformis fossa (FAN) and greater trochanter (TAN). These nails were compared experimentally for their fixation stability, proximal load transmission, and failure strength in an unstable intertrochanteric fracture model in cadaveric femurs. Vertical and axial loads were first applied to the intact femurs. Fractures were created, subsequent fixation applied, and the femurs underwent a series of both vertical and axial loading tests. There was no significant difference in strain readings between the nails for either axial loading or cyclical loading. There was no statistically significant difference between the loads to failure for the trochanteric nails and the standard antegrade nails. The average ultimate loadfor the FAN and TAN nails were 3010 N and 2830 N respectively. These two nails performed very similarly throughout our testing.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号