首页 | 本学科首页   官方微博 | 高级检索  
     


A method for in situ measurement of ionic concentration in picoliter samples of renal tubular fluid
Authors:P. Hansell  J. Sadowski  M. Sjöquist  H. R. Ulfendahl
Affiliation:(1) Department of Physiology and Medical Biophysics, University of Uppsala, Biomedical Center, P.O. Box 572, S-75123 Uppsala, Sweden;(2) Department of Applied Physiology, Polish Academy of Sciences, Medical Research Center, Warzaw, Poland
Abstract:A method permitting rapid and frequent determination of the total ionic concentration in renal tubular fluid samples, by measuring sample electrical impedance, is described. No special electrodes or sophisticated equipment are required. Since picoliter volumes are sufficient for measurement, tubular fluid dynamics are not significantly disturbed by the sampling procedure. The sample is aspirated into the tip of the standard glass micropipette and its ionic concentration, in the range of 100–1000 mmol/l, can be determined by any sensitive impedance-measuring device. In the present studies the Wheatstone bridge of the Wiederhielm apparatus for pressure measurement in micropuncture studies was used for the purpose. The obvious limitation of the method is that the measurements do not include nonelectrolytes and cannot differentiate between individual ion species.
Keywords:Micropuncture  microelectrode  microanalysis  ionic concentration
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号