首页 | 本学科首页   官方微博 | 高级检索  
     


Improved homogeneity of the transmit field by simultaneous transmission with phased array and volume coil
Authors:Nikolai I. Avdievich PhD  Sukhoon Oh PhD  Hoby P. Hetherington PhD  Christopher M. Collins PhD
Affiliation:1. Department of Neurosurgery, Yale University, New Haven, CT, USA;2. Center for NMR Research, Department of Radiology, Pennsylvania State University College of Medicine, Hershey, PA, USA
Abstract:

Purpose:

To improve the homogeneity of transmit volume coils at high magnetic fields (≥4 T). Due to radiofrequency (RF) field/tissue interactions at high fields, 4 T to 8 T, the transmit profile from head‐sized volume coils shows a distinctive pattern with relatively strong RF magnetic field B1 in the center of the brain.

Materials and Methods:

In contrast to conventional volume coils at high field strengths, surface coil phased arrays can provide increased RF field strength peripherally. In theory, simultaneous transmission from these two devices could produce a more homogeneous transmission field. To minimize interactions between the phased array and the volume coil, counter rotating current (CRC) surface coils consisting of two parallel rings carrying opposite currents were used for the phased array.

Results:

Numerical simulations and experimental data demonstrate that substantial improvements in transmit field homogeneity can be obtained.

Conclusion:

We have demonstrated the feasibility of using simultaneous transmission with human head‐sized volume coils and CRC phased arrays to improve homogeneity of the transmit RF B1 field for high‐field MRI systems. J. Magn. Reson. Imaging 2010;32:476–481. © 2010 Wiley‐Liss, Inc.
Keywords:TEM  high field MRI  CRC surface coil  RF field inhomogeneity  phased array
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号