Effect of HbS in the determination of HbA(2) with the TOSOH HLC-723G7 analyzer and the HELENA Beta-Thal Quik column kit |
| |
Authors: | Kalleas Christoforos Tentes Ioannis Margaritis Dimitrios Anagnostopoulos Kostas Toli Athina Pendilas Dimitrios Bourikas Georgios Tsatalas Constantinos Kortsaris Alexandros H |
| |
Affiliation: | Haematology Laboratory, General Hospital of Poligiros, Poligiros, Greece. |
| |
Abstract: | OBJECTIVES: The analytical performance of the TOSOH HLC-723G7 hemoglobin HPLC analyzer and the effect of the presence of HbS in the determination of HbA(2) using HPLC and manual column methods. DESIGN AND METHODS: The performance characteristics of the TOSOH HLC-723G7 analyzer in the determination of HbA(2) were compared to those of the HELENA Beta-Thal Quik column. The effect of HbS presence in the samples was quantified using the HELENA SAS-MX alkaline gel electrophoresis kit as the reference method. RESULTS: Within-run and between-run CVs for HbA(2) were better for the TOSOH HPLC analyzer than for the HELENA manual column method. The presence of HbS in the samples produces a strong positive bias in the %HbA(2) values when using both the HPLC and manual column methods, compared to the alkaline electrophoresis gel. CONCLUSION: Both the TOSOH HPLC and the manual column are reliable methods for %HbA(2) determination when no HbS is detectable in the samples. When HbS is present, the gel electrophoresis method gives more accurate results. |
| |