In vitro testing of a novel limb salvage prosthesis for the distal femur
Authors:Luca Cristofolini  Stefano Bini  Aldo Toni
Institution:

a Laboratory for Biomaterials Technology, Rizzoli Orthopaedic Institutes, Via di Barbiano 1/10, 40136, Bologna, Italy

b Mechanical Engineering Department, DIEM, University of Bologna, Bologna, Italy

c Department of Orthopaedic Surgery, Kaiser Permanente Medical Center, Redwood City, CA, USA

d Orthopaedics Clinic, University of Bologna, Bologna, Italy

Abstract:Objective. The aim of this paper was to define strain pattern in the host bone following distal femoral resection and implantation of a massive prosthesis. Two methods of coupling the prosthesis to the bone were compared: the Compliant Pre-Stress device, and a standard cemented tumour prosthesis.

Design. The composite femur model was selected to minimize variables. Four femurs were tested before and after implantation. Both coaxial and cantilever loading were applied.

Background. Cemented distal femoral replacement following resection of malignant tumours has a high failure rate at 5 years and is associated with extensive bone resorption thought to be secondary to stress shielding.

Methods. Strain was measured in the medial and lateral sides at four levels with physiologic loads applied, in the intact, Compliant Pre-Stress, and cemented femurs. Repeated measurements were taken. Strains in the implanted femur were calculated as percentage of the intact, and statistically analyzed.

Results. The most reproducible results were noted in cantilever bending (variability <5%). The Compliant Pre-Stress device demonstrated a more physiologic strain pattern than the cemented stem. The most significant difference between the two implants was in the area adjacent to the interface.

Conclusion. The Compliant Pre-Stress device shows less stress shielding than a standard cemented implant. The protocol described and the use of composite femurs demonstrated reproducible results.

Keywords:Author Keywords: Arthroplasty  Massive prosthesis  Femur  Strain gauge  Stress shielding  Resection