Acute and chronic stability of atrial screw-in leads
Authors:Y Nakazato  Y Nakata  Y Ohno  M Sumiyoshi  H Hisaoka  S Ogura  H Yamaguchi  N Hazato
Institution:Department of Cardiology, Juntendo University School of Medicine, Tokyo, Japan.
Abstract:The purpose of this study was to evaluate the acute and chronic stability of the atrial screw-in lead. In this study, we used CPI model 4165, 4166, and 4266 porous tip screw-in leads in 32 patients (12 for AAI pacing and 20 for DDD pacing). All of these leads were fixed to the free wall of the right atrium and used as the unipolar type. Acute voltage and current thresholds, lead impedance, P-wave amplitude, and chronic threshold were measured. Early and late complications were also investigated. The average acute stimulation thresholds at 0.6 msec pulse width were 0.82 +/- 0.43 V and 1.1 +/- 0.6 mA. Mean lead impedance was 627.2 +/- 83.1 ohms, and mean P wave amplitude was 3.1 +/- 1.1 mV. After an average follow-up period of 32 months (range: 2-72 months), we found the results of the chronic threshold study to be satisfactory. The thresholds were usually below 0.1 msec pulse width with the nominally programmed pacemaker output. Only one patient required a higher output due to an increased threshold. With regard to complications, neither lead dislodgment nor cardiac perforation was observed. In conclusion, acute and chronic thresholds were satisfactory and no serious complications occurred. Therefore the atrial screw-in lead has long-term reliability and stability.