首页 | 本学科首页   官方微博 | 高级检索  
检索        

动态轮廓眼压计与非接触眼压计在正常青年人坐位眼压昼夜波动趋势的研究
引用本文:曹瑞丹,张越鹏,耿佳,圣娟娟,张作明.动态轮廓眼压计与非接触眼压计在正常青年人坐位眼压昼夜波动趋势的研究[J].国际眼科杂志,2009,9(3):498-501.
作者姓名:曹瑞丹  张越鹏  耿佳  圣娟娟  张作明
作者单位:第四军医大学航空航天医学系航临教研室,中国陕西省西安市,710032
摘    要:目的:采用PASCAL动态轮廓眼压计(dynamic contour to-nometer,DCT)及KEELER非接触眼压计(non contact to-nometer,NCT)对一组健康青年昼夜眼压进行监测,描述其波动规律,并比较两种方法异同。方法:以PASCAL动态轮廓眼压计对一组19~22岁健康青年27例(54眼)进行24h眼压监测;以KEELER非接触眼压计对一组19~22岁健康青年84例(168眼)进行24h眼压监测。测量自早晨08∶00开始至次日凌晨06∶00,每隔2h为一个观察时间点,每个时间点测量3次,以平均值作为该时间点眼压。结果:DCT法与NCT法所测得的眼压平均值分别为15.76±2.37mmHg和13.77±4.33mmHg,DCT组眼压均值的最高值出现在08∶00,最低值出现在20∶00;NCT组眼压均值的最高值出现在06∶00,最低值出现在22∶00。DCT组双眼眼压波动值分别为:左眼4.56±1.19mmHg,右眼4.12±1.21mmHg;NCT组双眼眼压波动值分别为:左眼4.71±2.23mmHg,右眼5.00±1.50mmHg,两种方法左右眼间眼压值差异无统计学意义(F=0.668,P=0.415),男女眼压值差异无统计学意义(F=1.459,P=0.229),12个时间点之间眼压值差异有统计学意义(F=64.407,P=0.000),且变化趋势呈为二次抛物线,两种方法随时间变化趋势有统计学意义(F=16.615,P=0.000)。结论:正常青年眼压呈昼夜节律变化,于白昼结束时最低,夜晚结束时最高。不同的测量方法在各时间点所得测量值及昼夜变化趋势上存在差异,且采用动态轮廓眼压计在多个时间点上均比采用非接触眼压计所测值高。临床上在应用眼压作为原发性青光眼诊断和疗效判断时,应注意测量时间及方法选择上的差别。

关 键 词:动态轮廓眼压计  非接触眼压计  青光眼

Research on dynamic contour tonometer and non-contact tonometer intraocular pressure in normal young people sitting day and night fluctuations trend
Rui-Dan Cao,Yue-Peng Zhang,Jia Geng,Juan-Juan Sheng,Zuo-Ming Zhang.Research on dynamic contour tonometer and non-contact tonometer intraocular pressure in normal young people sitting day and night fluctuations trend[J].International Journal of Ophthalmology,2009,9(3):498-501.
Authors:Rui-Dan Cao  Yue-Peng Zhang  Jia Geng  Juan-Juan Sheng  Zuo-Ming Zhang
Institution:Rui-Dan Cao,Yue-Peng Zhang,Jia Geng,Juan-Juan Sheng,Zuo-Ming Zhang Air Pro Teaching and Research,Department of Aerospace Medicine,the Fourth Military Medical University,Xi'an 710032,Shaanxi Province,China
Abstract:AIM:To monitor intraocular pressure(IOP)fluctuation change in healthy young with PASCAL dynamic contour tonometer(DCT)and KEELER non-contact tonometer(NCT)and to compare the measuring results with the two methods..METHODS:Prospective study including 54 eyes of 27 healthy young with DCT and another 168 eyes of 84 healthy young with NCT for 24 hours.All the healthy young are aged from 19 to 22,and every two hours should be a time interval.IOP was measured from 08:00 am to next day 06:00am.The measurement at e...
Keywords:dynamic contour tonometer  non-contact tonometer  glaucoma  
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号