首页 | 本学科首页   官方微博 | 高级检索  
检索        


Assessment of coronary flow reserve with transthoracic Doppler echocardiography: comparison with intracoronary Doppler method
Authors:Kim S M  Shim W J  Lim H E  Hwang G S  Song W H  Lim D S  Kim Y H  Seo H S  Oh D J  Ro Y M
Institution:Department of Internal Medicine, Korea University College of Medicine, Seoul.
Abstract:To evaluate the feasibility and usefulness of transthoracic Doppler echocardiography (TTDE) as a non-invasive method in recording distal anterior descending (LAD) coronary flow velocity, we compared coronary flow reserve (CFR) measured by TTDE with measurements by intracoronary Doppler wire (ICDW). Twenty-one patients without LAD stenosis were studied. ICDW performed at baseline and after intracoronary injection of 18 microg adenosine. TTDE was performed at baseline and after intravenous adenosine (140 microg/kgmin for 2 min). Adequate Doppler recordings of coronary flow velocities during systole were obtained in 14 of 21 study patients (67%) and during diastole in 17 (81%) patients. Baseline and hyperemic peak diastolic flow velocities measured by TTDE were significantly smaller than those obtained by ICDW (p<0.05). However, diminishing trends of diastolic and systolic velocity ratio after hyperemia were similarly observed in both methods. CFR obtained by TTDE (3.0+/-0.5), was higher than the value calculated by ICDW (2.5+/-0.4). There were significant correlations between the values obtained by the two methods (r=0.72, p<0.01). It is concluded that TTDE is a feasible method in measuring coronary flow velocity and appears to be a promising non-invasive method in evaluating CFR.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏