首页 | 本学科首页   官方微博 | 高级检索  
检索        


Coarctation of the aorta treated with the Advanta V12 large diameter stent: Acute results
Authors:Elchanan Bruckheimer MBBS  Einat Birk MD  Raul Santiago MD  Tamir Dagan MD  Carlos Esteves MD  Carlos AC Pedra MD
Institution:1. Section of Pediatric Cardiology, Schneider Children's Medical Center—Israel, Petach Tikva, Israel;2. Section of Pediatric Cardiology, Dante Pazzanese, Sao Paulo, Brazil
Abstract:Objectives: To report on the early results of treatment of coarctation of the aorta by dilation with a new polytetrafluoroethylene covered stent. Background: Transcatheter dilation of aortic coarctation carries the risk of aneurysm or rupture. Covered stent implantation reduces this risk but requires a large delivery system. The Advanta V12 LD covered stent is premounted and requires a 9–11 Fr delivery system. Methods: Covered stents on balloons of a diameter sufficient to anchor the stent in the coarctation were implanted using the smallest available delivery system. Secondary dilation with larger diameter balloons was performed until the pressure gradient was <20 mm Hg and the stent was opposed to the aortic wall. Results: Twenty‐five patients with aortic coarctation underwent stent implantation. Coarctation diameter increased from (6.3 ± 3.5) mm to (14.4 ± 2.3) mm (P < 0.0001). Peak pressure gradient decreased from (25.3 ± 11.6) mm Hg to (2.5 ± 3.0) mm Hg (P < 0.0001). The stent achieved the desired diameter in all cases. There were no complications. At short‐term median follow‐up of 4.9 months, all patients are alive and well with no evidence of recoarctation or aneurysm. Conclusions: These initial results show that the covered Advanta V12LD stent is safe and effective in the immediate treatment of coarctation of the aorta through a low profile delivery system of 8–11 Fr. Long term follow up is required. © 2009 Wiley‐Liss, Inc.
Keywords:heart defects  congenital  catheterization  radiology interventional
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号