Mechanical valve dysfunction after percutaneous perimitral leak closure: Salvage by percutaneous occluder retrieval
Authors:Emmanouil S. Brilakis MD  PhD   FSCAI  Laura J. Collins MD  Owen Obel MD  Subhash Banerjee MD  FSCAI
Affiliation:VA North Texas Healthcare System and University of Texas Southwestern Medical Center, Dallas, Texas
Abstract:Percutaneous closure of perivalvular regurgitation may be complicated by prosthetic valve impingement and dysfunction. We describe a patient with two perimitral defects who developed severe mitral stenosis due to mitral leaflet impingement after percutaneous implantation of three amplatzer muscular ventricular septal defect occluder devices. The largest amplatzer occluder device was successfully snared with restoration of normal prosthetic valve motion. © 2010 Wiley‐Liss, Inc.
Keywords:mechanical valves  mitral regurgitation  occluder devices