首页 | 本学科首页   官方微博 | 高级检索  
     


MR-compatible treadmill for exercise stress cardiac magnetic resonance imaging
Authors:Foster Eric L  Arnold John W  Jekic Mihaela  Bender Jacob A  Balasubramanian Vijay  Thavendiranathan Paaladinesh  Dickerson Jennifer A  Raman Subha V  Simonetti Orlando P
Affiliation:Dorothy M. Davis Heart and Lung Research Institute, The Ohio State University, Columbus, Ohio, USA.
Abstract:
This article describes an MR-safe treadmill that enables cardiovascular exercise stress testing adjacent to the MRI system, facilitating cardiac MR imaging immediately following exercise stress. The treadmill was constructed of nonferromagnetic components utilizing a hydraulic power system. Computer control ensured precise execution of the standard Bruce treadmill protocol commonly used for cardiovascular exercise stress testing. The treadmill demonstrated no evidence of ferromagnetic attraction and did not affect image quality. Treadmill performance met design specifications both inside and outside the MRI environment. Ten healthy volunteers performed the Bruce protocol with the treadmill positioned adjacent to the MRI table. Upon reaching peak stress (98 ± 8% of age-predicted maximum heart rate), the subjects lay down directly on the MRI table, a cardiac array coil was placed, an intravenous line connected, and stress cine and perfusion imaging performed. Cine imaging commenced on average within 24 ± 4 s and was completed within 40 ± 7 s of the end of exercise. Subject heart rates were 86 ± 9% of age-predicted maximum heart rate at the start of imaging and 81 ± 9% of age-predicted maximum heart rate upon completion of cine imaging. The MRI-compatible treadmill was shown to operate safely and effectively in the MRI environment.
Keywords:MRI  heart  cardiac function  stress test
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号