Confirmation of high blood flow rates through 150 μ filter/high-flow tubing
Authors:Kenneth V. Iserson MD   FACEP  Mary Ann Knauf MT
Affiliation:

a Section of Emergency Medicine, Department of Surgery, University of Arizona College of Medicine, Tucson, Arizona, USA

b Blood Bank, University of Arizona College of Medicine, Tucson, Arizona, USA

Abstract:Many new products designed to assist in rapid blood infusion are appearing. Some highly touted and routinely used devices for intravenous (IV) infusion have recently been shown to be, at least in part, defective. A tubing with an in-line 150 μ filter (150 μ High-Flow Blood Filter; Saftifllter® Blood Administration Sets; Cutter Biological, Berkeley, CA 94710) has recently been introduced to facilitate rapid blood transfusion. It is claimed that at least 8.5 units of blood can be rapidly run through each set before replacement is necessary. To test this under simulated clinical conditions, four sets of ten random units of outdated erythrocytes at 4 to 9°C were each admixed with 250 mL 70°C 0.9 NaCl and infused through the system under a constant 300 mmHg pressure. Two sets infused through unmodified tubing flowed at an average of 25 mL/sec (1500 mL/min) before there was an appreciable slowing of the flow rate. Two sets with 8 Fr catheters attached infused at an average of 22 mL/sec (1320 mL/min) before there was an appreciable slowing of the flow rate. Even after the flow slowed, the 9th and 10th units infused at an average greater than 10 mL/sec (600 ml/min). The tubing/filter exceeded the manufacturer's published claims. This tubing/filter appears to be one element that could be an effective component of a high-flow infusion system.
Keywords:Author Keywords: intravenous   tubing   large bore   filters   rapid fluid infusions   fluid therapy
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号