The measurement of anterior chamber depth andaxial length with the IOLMaster compared withcontact ultrasonic axial scan |
| |
Authors: | Quan-Hao Bai Jun-Li Wang Qing-Qiang Wang Qi-Chang Yan Jin-Song Zhang |
| |
Affiliation: | Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang 110005, Liaoning Province, China;Health Management and Education School, Capital Medical University, Beijing 100069, China;Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang 110005, Liaoning Province, China;Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang 110005, Liaoning Province, China;Department of Ophthalmology, the Fourth Affiliated Hospital of China Medical University, Shenyang 110005, Liaoning Province, China |
| |
Abstract: | AIM: To compare the measurement of anterior chamber depth (ACD) and axial length (AL) by IOLMaster and contact ultrasonic (US) axial scan (A-scan).· METHODS: Measurements of ACD and AL were prospectively obtained in 137 eyes of 121 subjects with the IOLMaster compared with measurements with the US.· RESULTS: There was an excellent correlation between IOL Master and US measurements for the ACD (r =0.823;P < 0.001) and AL (r =0.996;P <0.001). The mean values of the parameters measured by IOLMaster and US were, respectively, as follows: ACD, 2.94±0.49mm, 2.58±0.51mm; AL, 24.37±3.04mm, 23.81±2.83mm. The mean differences of ACD and AL values between IOLMaster and US measurements were 0.36±0.30mm, 0.56±0.34 mm respectively, and they proved to be statistically significant (P < 0.001), with the 95% limits of agreement (LoA) from -0.08mm to +0.38mm for ACD and from -0.09mm to +0.69mm for AL.· CONCLUSION: As noncontact biometry, IOLMaster provides accurate values. A high degree of agreement between US and IOLMaster was noted. It not only has the advantage of performing noncontact examinations, but also produces various additional data simultaneously and may thus obviate the need for multiple examinations. Further studies are needed to assess the interchangeability of measurements in clinical practice. |
| |
Keywords: | anterior chamber depth axial length IOLMaster contact ultrasonic axial scan |
|
| 点击此处可从《国际眼科》浏览原始摘要信息 |
|
点击此处可从《国际眼科》下载免费的PDF全文 |
|