Advanced multiple-beam equalization radiography in chest radiology: a simulated nodule detection study
Authors:Kool  LJ; Busscher  DL; Vlasbloem  H; Hermans  J; vd Merwe  PC; Algra  PR; Herstel  W
Institution:Department of Diagnostic Radiology, University Hospital Leiden, The Netherlands.
Abstract:To evaluate the efficacy of AMBER, a multiple-beam equalization system for chest radiography, the authors performed a nodule detection study using an anthropomorphic chest phantom. AMBER and conventional images were compared. The images were read by four observers, and analysis was done by means of modified receiver-operating characteristic (ROC) curves (free ROC curves FROC]). The results of the FROC analysis show a significant increase in the detectability of nodules (P less than .001) projected over the mediastinum with the use of AMBER. No significant difference between AMBER and conventional images was noted in detectability of nodules projected over the lung.