首页 | 本学科首页   官方微博 | 高级检索  
     


Photostimulable storage phosphor image acquisition: Evaluation of three commercially available state-of-the-art systems
Authors:Jonathan E. Tucker  Maricela Contreras  Ronald J. Wider  Martin G. Radvany  Anna K. Chacko  Rashmikant B. Shah
Affiliation:1.Department of Radiology,Brooke Army Medical Center,Fort Sam Houston
Abstract:Photostimulable storage phosphor (PSP) image acquisition systems have been available for several years. The technology has had the opportunity to mature; however, there has not been an independent comparison of recently marketed commercial systems. For this study, three computed radiography (CR) systems using PSP technology (Kodak CR System 400 with autoloader [Eastman Kodak, Rochester, NY], Fuji FCR AC-3CS [Fuji Medical Systems, Stamford, CT], and Agfa ADC Compact [Bayer Corp, Ridgefield Park, NJ]) were connected to an IBM RadWorks diagnostic radiology workstation (IBM Corp, White Plains NY) and evaluated for conformance to their performance specifications using guidance provided in the most recent draft acceptance testing protocol from Task Group No. 10, American Association of Physicists in Medicine. In addition, the physical requirements (eg, space and power) and connectivity to another manufacturer’s diagnostic workstation were examined. X-ray technologist comfort with each PSP imaging system and an assessment by our supporting biomedical equipment maintenance activity of their ability to service each PSP imaging system were also considered.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号